![SciCan STATMATIC smart 1 Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/scican/statmatic-smart-1/statmatic-smart-1_instructions-for-use-manual_1217167010.webp)
2 Commissioning
Page 10
STAT
MATIC
smart 1,2,3
EN
Commissioning
Turn on the mains switch.
Select the desired language with the arrow keys next to the display (Page 10).
Set the date and time (Page 11).
"WARNING Fill with dental oil" and/or "WARNING Fill with cleaner" will be displayed.
Pull off top cover with both hands approx. 15 mm (overcome magnetic resistance) and pull off upwards.
Refill the media containers (also see chapter 4 "Maintenance").
Observe the color code on the media container!
Blue = dental oil / green = cleaner
Only use original SciCan "Dental Oil" and SciCan "Cleaner".
Use of other media may put patients and users at risk and damage the device and the instruments.
Turn the metal flaps to the back in order to open the media container.
Fill up the container to the filling level marking. Upon exceeding the minimum filling level, a 1-second beep will
sound. If the container is full, a 2-second beep will sound 2 times. The beep is always followed by a visual sign
on the container.
Close the media container.
Attach the top cover and close until it stops.
Turn the device off using the power switch and then back on again.
The message "Fill media system" will be displayed after filling the dental oil and cleaner containers. The device
will automatically start this function and run for about 3 minutes.
Filling the media system
This is only necessary if the function did not start automatically or if a container has been completely empty for more
than 3 days. The containers must be filled already for this function.
Turn on the device
Press the
button. "Export log data" will appear on the display.
Press the
button once until "Fill media system" is highlighted in black on the display.
Using the
button, the "Fill media system" menu is selected.
The media system will be filled automatically. During this time, a clock will run backwards until 00:00.
Press the
button to return to the "Cl Maintenance" main menu
Setting the language
Turn on the device
Press the
button. "Export log data" will appear on the display.
Press the
button 4x until "Language setting" is highlighted in black on the display.
The "Language setting" menu is selected using the
button.
The desired language is selected using the
button.
Save the language selection using the
button.
Return to the "Language setting" menu using the
button.
Press the
button to return to the "Cl Maintenance" main menu.