TENS eco 2

TRANSCUTANEOUS 

NERVE AND MUSCLE STIMULATOR

Instructions for Use  Art.-No. 101446-V05   

Содержание TENS eco 2

Страница 1: ...HOLISTIC HEALTH EN TENS eco 2 TRANSCUTANEOUS NERVE AND MUSCLE STIMULATOR Instructions for Use Art No 101446 V05 ...

Страница 2: ...n P10 P11 P22 P23 16 Technical information 17 Description of the symbols 17 Technical specification 17 Pulse form 18 Maintenance and cleaning 18 Warranty 18 Guarantee 18 Combination 18 Classification 18 Reference sheet for battery retraction and disposal 19 Delivery content 19 Accessories 20 Stimex self adhesive electrodes 20 Silicone Electrode 20 Stimex garment electrodes 21 Other Accessories 21 ...

Страница 3: ...on Pain can indicate serious disorders in the body and must be examined by a physician Even if the application of TENStem eco basic is successful and leads to a significant pain relief this is not to be equated with a cure of the cause of pain Contraindications In which cases shall a patient not use the device or if only with consultation of a medical practitioner Patients with implanted electroni...

Страница 4: ...ctrodes placed on the chest and the upper back or crossing over the heart Warning Application of electrodes near the thorax may increase the risk of cardiac fibrillation In case of electrode placement in the thorax area intensive high frequency stimulation above approx 15 Hz may lead to respiratory disorders during stimulation 3 Keep the product away from water and other liquids as this may cause ...

Страница 5: ... burns beneath the electrodes and it may damage to the product 12 Caution Attach the electrodes to the skin so that the electrode sur face has even and complete skin contact Additionally make sure that the electrodes are placed at least 2 cm apart from each other Other wise high current densities may occur at certain points on the skin re sulting in painful skin lesions 13 Before applying the elec...

Страница 6: ...of sore muscle can occur Shorten the treatment time and choose a low intensity of stimulation during the first treatment sessions to avoid pain aggravation Muscle contraction During high frequency stimulation above approx 15 Hz persistent muscle contractions can occur Electrode placement in the thorax area may lead to respiratory troubles during the stimulation If the applied area is near or on th...

Страница 7: ...lue Back pain Shoulder pain one channel two channels 1 1 2 2 Shoulder neck pain Epicondylitis radialis Epicondylitis ulnaris Arthrosis hip Kaada stimulation one channel only right handed left handed Examples of electrode placement ...

Страница 8: ...structions for Use TENS eco 2 8 Sciatic pain Gonarthrosis knee pain Ankle joint pain Trigeminus neuralgia Migraine Tension headache Carpal tunnel syndrome Polyarthritis Pain of achilles tendon heel pain Polyneuropathy ...

Страница 9: ...s two channels with stimulation gloves two channels Attention Stimulation socks and gloves have to be moistured well Amputation pain stomp pain Amputation pain phantom pain Postzoster neuralgia contralateral Channel 1 Channel 2 Channel 1 Channel 2 ...

Страница 10: ...for almost all kinds of pain program 3 s 100 Hz then 3 s 2 Hz etc 100 2 150 200 30 P 7 Burst Alternative to program 3 more comfortable 100 Hz for 0 25 s 100 150 30 P 8 Modula tion Alternative to the other programs in case of therapy resistance Automatic frequency run from 2 Hz to 80 Hz within 7 5 s 2 80 200 100 30 P 9 Muscle stimula tion Atrophy prophylaxis muscle strengthening Increase 2 s workin...

Страница 11: ...re quency 1 120 50 400 1 99 U 7 Burst Alternative to pro gram no 3 more comfortable 0 25 s long impulse with a frequency to be set 50 120 50 400 1 99 U 8 Modula tion Alternative to other programs if resis tant to therapy Automatic frequency run from one frequency to ano ther within 7 5 s First fre quency 1 50 Last fre quency 80 First pulse width 100 400 1 99 U 9 Muscle Stimula tion Prophylaxis tre...

Страница 12: ...y of new TENS eco 2 units need to be charged before first use Please read the chapter charging the battery on page 10 Switch the commutator no 8 to the ON position Switch on the TENS unit by pressing the key no 5 The program shown on the display is the last which was used during the previous session If the display shows a key symbol on the upper left side the TENS eco 2 has previously interlocked ...

Страница 13: ...ically If you press the key the stimulation stops and the TENS eco 2 switches OFF Interlocking the keypad Choose the desired program by pressing the key see chapter choosing a program Press and hold for 3 seconds the key and the right keys to interlock the TENS unit on the chosen program The TENS eco 2 can now only be used with the chosen program A key symbol appears on the display adjacent to the...

Страница 14: ...de on the charger illuminates in red Charge the TENS unit until the diode on the charger illuminates in green The battery is fully charged if the light on the charger appears in green Plug OFF the charging cable from the mains circuit Caution Do not charge the battery for more than 3 hours Important Do not forget to switch the commutator ON again after charging Socket Commutator Setting date and t...

Страница 15: ... 5 fifth day Pulse width 200 μs Therapy session no 3 You can save up to 90 therapy sessions respectively the last 3 of each day With the left key you can return to the values of the previous day until you reach the first recorded day The left key lets you proceed to the next day until you reach the current or last day The right key brings you to the parameters of the previous therapy session of th...

Страница 16: ...treatment method also has an enhanced effect on lymphatic drainage stimulation The placement of electrodes for dynamic stimulation depends on the direction of stimulation rising falling stimulation or sideways stimulation Rising or falling Sideways Rising Falling stimulation stimulation wave wave for back or legs for gluteus or stomach muscle Kanal 1 Kanal 2 Kanal 1 Kanal 2 Channel 1 Channel 2 Kan...

Страница 17: ...ng By doing this you will help to protect the environment Manufacturer Date of manufacture Attention The safe use of the product requires following the instructions for use Keep dry Technical information Description of the symbols Technical specification Two channel nerve and muscle stimulator with electrically insulated channels constant current characteristic output short circuit control element...

Страница 18: ...Legal right of warranty is applied according to German Civil Code Guarantee The manufacturer issues a guarantee of 12 months from the date of purchase The guarantee does not apply in the following cases damage due to improper handling defects the customer is aware of on the date of purchase damage caused by the customer for wearing parts and consumable supplies like for instance electrodes batteri...

Страница 19: ...e We are also obliged to take back used batteries whereat our obligation restricts to those which are or were part of our product range Therefore you may send back used batteries to us adequately stamped by mail or render these at our warehouse free of charge at the following address schwa medico GmbH Wetzlarer Str 41 43 35630 Ehringshausen Germany Please refer to the following figures to see the ...

Страница 20: ...mm 4 283600 Stimex 50x90 mm 2 283000 Stimex 50x130 mm 2 283100 Stimex 80x130 mm 2 281027 Stimex sensitive 50x50 mm 4 Silicone Electrode Apply electrode gel to the flat side of the electrodes and attach them to the skin with a strip of Silk Tape Do not stick to open wounds The conductivity of the electrodes slowly decreases after 6 months Replace after approx 12 months of intensive use Please clean...

Страница 21: ...ng the therapy Do not wear any metallic jewellery or watches during therapy For further safety information and for detailled cleaning measures please read the instructions for use of the Stimex garment electrodes see below Gloves Art No Article Size Pair 107014 Stimulation gloves S 1 107021 Stimulation gloves M 1 107022 Stimulation gloves L 1 451600 0438 Instructions for use Socks Art No Article S...

Страница 22: ... or empty Charge the rechargeable battery If charging is not successful replace the rechargeable battery If a new rechargeable battery does not work either use a new charging unit The unit suddenly switches the intensity back to 0 One or both electrodes have slipped or do not stick properly to the skin a Check that the electrodes are properly attached and reattache them if necessary b If necessary...

Страница 23: ...English 23 ...

Страница 24: ... Am Geiersberg 6 I 35630 Ehringshausen I Deutschland www pierenkemper eu EXPORT DEPARTMENT Wetzlarer Strasse 41 43 I 35630 Ehringshausen I Germany Tel 49 6443 8333 113 I Fax 49 6443 8333 119 Email export schwa medico de I www schwa medico com HOLISTIC HEALTH ...