GB
57
9. Accessories: foot switch
and brightfield-dark-
field adapter kit
The foot switch can be connected to the MC 1500 connection (15).
When repeatedly pressed down it calls up the M1 to M4 memory stations and
makes it possible to quickly switch over the stored illumination conditions
without having to press a memory key.
(M1
➔
M2
➔
M3
➔
M4
➔
M1
➔
...)
The brightfield-darkfield adapter kit makes it possible to connect one of the
LED brightfield ringlights (S80-25, S80-55) with the darkfield ringlight
(S40-10D) to a stereo microscope.
If the ringlights combined on the microscope are then connected to the
MC 1500 one gets an incident light illumination that makes it possible to
quickly switch between brightfield and darkfield. In addition this facilitates to
comfortably generate mixed light. When the adapter kit was developed great
care was taken to minimize the occurrence of scatter light and ambient light on
the specimen.
The brightfield-darkfield adapter kit is made up of three components: the
connection adapter, and two distance rings.
To set up the adapter kit first please screw the connection adapter into the thread
of the S40-10D darkfield ringlight. Then fix the connection adapter to a VisiLED
brightfield ringlight by sliding the adapter over the ringlight and fixing with
three screws.
The brightfield-darkfield illumination is assembled on the microscope by sliding
the brightfield ringlight over the microscope objective and screwing tight.
The acceptance diameter of the ringlights is 66 mm. To fix onto microscope
objectives with other diameters please use intermediate rings (accessory).
If the darkfield ringlight is not situated in the recommended 5 mm to 15 mm
working distance above the microscope specimen, please carry out the assem-
bling of the adapter kit as follows:
First screw one or two of the supplied distance rings into the VisiLED brightfield
ringlight. Then slide the connection adapter with the darkfield ringlight over the
distance ring(s) on the brightfield ringlight and fix with the three screws.
Foot switch
Brightfield-darkfield adapter kit
Содержание VisiLED MC 1000
Страница 1: ...VisiLED Series Gebrauchsanweisung DE GB FR ...
Страница 2: ......
Страница 4: ......
Страница 9: ...153 MC 1500 A B 110 110 40 A B RS232 21V dc I O 153 40 DE 7 2 Controller 2 1 MC 1500 in mm ...
Страница 41: ...GB 39 153 MC 1500 A B 110 110 40 A B RS232 21V dc I O 153 40 2 Controller 2 1 MC 1500 in mm ...
Страница 102: ...100 Notes ...
Страница 103: ...101 ...