GB
45
•
Please take care that the VisiLED system is used with electronic power supplies
and power cords specified by SCHOTT – only then is faultless operation and
conformity with standards guaranteed!
•
Please ensure that the electronic power supply is only operated with the
specified mains voltage. The power supply must only be plugged into sockets
which are earthed. More details regarding the electronic power supply can be
found in Section 12, Technical data.
•
In the event of insufficient cooling the integrated thermo guard switches
off the illumination unit temporarily to protect the LEDs.
•
The illumination system has been developed for operation in dry rooms only.
•
The system must not be used in explosive areas.
•
Safe disconnection from the power supply takes places only by pulling out the
mains plug.
•
Before changing the system configuration always switch off the controller and
pull out the mains plug.
•
For reasons of energy saving we recommend to pull out the mains plug and
so disconnect the electronic power supply from the mains voltage when the
VisiLED system is not in use.
•
The electronic power supply or the VisiLED controllers or illumination units
must not be opened or dismantled. Technical modifications to any parts of the
VisiLED system are forbidden. Repairs must be carried out only by the manu-
facturer or by its authorized customer service agencies.
•
Please ensure that every user of the VisiLED system has quick access to these
instructions.
•
The manufacturer is not liable for damage caused by failure to obey these
instructions.
•
The LED transmitted light brightfield stage has been developed for operation
in a microscope stand or in another metallic heat sink.
•
If the transmitted light brightfield stage is operated without heat sink the
metal housing can heat up to approx. 45°C. However, the diffuser disc and the
specimen field do not heat up to the same level.
•
If the housing is accidentally touched the temperature may seem unpleasantly
warm but it is technically safe. The operation of the illumination is not
affected!
Information for operating the LED
transmitted light brightfield stage
Important information
Содержание VisiLED MC 1000
Страница 1: ...VisiLED Series Gebrauchsanweisung DE GB FR ...