7.3
Calibration
Why calibrate?
pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is dis-
played. Calibration determines the current values of the zero point and
slope of the electrode and stores them.
Thus, you should calibrate at regular intervals.
When do you have to
calibrate?
After connecting a non-ID electrode
After the CalClock has expired
Buffer sets for
calibration
You can use the buffer sets quoted in the table for an automatic calibra-
tion. The pH values are valid for the specified temperature values. The
temperature dependence of the pH values is taken into account during
the calibration.
No.
Buffer set *
pH values
at
1
DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1,679
4,006
6,865
9,180
12,454
25 °C
2
TEC
2,000
4,010
7,000
10,011
25 °C
3
Merck1
* (Technical buffers)
4,000
7,000
9,000
20°C
4
Merck2
*
1,000
6,000
8,000
13,000
20°C
5
Merck3
*
4,660
6,880
9,220
20°C
6
Merck4
*
2,000
4,000
7,000
10,000
20°C
7
Merck5
*
4,010
7,000
10,000
25 °C
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......