Legacy Protocol
Settings
D20980.026 Rev B
CV-LS Programming Guide
Page 26
4.1.4
Controls Lockout
4.1.4.1
Legacy
&K#
Gets/Sets which external inputs are enabled on the unit.
Input Options (#): 0 = No Lockouts, 1 = Front Lockout Only, 2 = Multiport Lockout Only,
3 = Both Front and Multiport Lockout, ? = Query
Return string:
&kX
Return value (
X
): 0 = No Lockouts, 1 = Front Lockout Only, 2 = Multiport Lockout Only,
3 = Both Front and Multiport Lockout
4.1.4.2
CV-LS
&HLF#
Gets/Sets the lockout status of the front switch and front knob on the unit.
Input Options (#): 0 = Disable, 1 = Enable, ? = Query
Return string:
&hlfX
Return value (
X
): 0 = Disable, 1 = Enable
Note:
Front Button will still respond to factory reset override, even when the front
controls are locked out.
&HLM#
Gets/Sets the lockout status of the analog controls on the Multiport.
Input Options (#): 0 = Disable, 1 = Enable, ? = Query
Return string:
&hlmX
Return value (
X
): 0 = Disable, 1 = Enable
Note:
Digital inputs will still respond allowing for remote switch controls of the unit.