Legacy Protocol
Controls
D20980.026 Rev B
CV-LS Programming Guide
Page 24
3.6
Fan Settings
3.6.1
Manual Fan Control
&GE#
Gets/Sets the fan override control. When the fan override is active, the fan will turn on
to the set speed regardless of the state of the unit. The fan will remain on for 15
seconds after the last fan setting change.
Input Options (#): 0 = Auto Mode, 1 = Manual Mode, ? = Query
Return string:
&geX
Return value (
X
): 0 = Auto Mode, 1 = Manual Mode
3.6.2
Fan Speed
&GS#
Gets/Sets the fan speed set point control.
Input Options (#): 0 = Auto Input Options (#): 0 – 1000 = Target Speed, ? = Query
Return string:
&gsX
Return value (
X
): 0 – 1000 = Target Speed