Legacy Protocol
Controls
D20980.026 Rev B
CV-LS Programming Guide
Page 20
3.3.5
Phase Shift
3.3.5.1
Legacy
&RP#
Gets/Sets the amount of offset from the internal trigger.
Input Options (#): 0 – 1000 = Phase Shift, ? = Query
Return string:
&rpX
Return value (
X
): 0 – 1000 = Phase Shift
3.3.5.2
CV-LS
&RP#,#
Gets/Sets the amount of offset per channel from the internal trigger.
First Input Option (#): 1 – 4 = Channel
Second Input Option (#): 0 – 1000 = Phase Shift, ? = Query
Return string:
&rpX,Y
Return value (
X
): 1 – 4 = Channel
Return value (
Y
): 0 – 1000 = Phase Shift
3.3.6
Waveform Polarity
&RJ#,#
Gets/Sets whether the active part of the waveform has the LED On or Off. The LED will
be on with an Active High setting, or Off with an Active Low setting, during the set Duty
Cycle %.
First Input Option (#): 1 – 4 = Channel
Second Input Option (#): 0 = Active Low, 1 = Active High, ? = Query
Return string:
&rjX,Y
Return value (
X
): 1 – 4 = Channel
Return value (
Y
): 0 = Active Low, 1 = Active High