Legacy Protocol
Controls
D20980.026 Rev B
CV-LS Programming Guide
Page 19
3.3
Continuous Strobe
3.3.1
Continuous Strobe Enable
&RM#
Gets/Sets if the continuous strobe mode is enabled.
Input Options (#): 0 = Disabled, 1 = Enabled, ? = Query
Return string:
&rmX
Return value (
X
): 0 = Disabled, 1 = Enabled
3.3.2
Single Channel Operation
&RB#
Gets/Sets if the unit will behave as a single channel device during continuous strobe.
Input Options (#): 0 = Quad Channel Mode, 1 = Single Channel Mode,
? = Query
Return string:
&rbX
Return value (
X
): 0 = Quad Channel Mode, 1 = Single Channel Mode
3.3.3
Frequency
&RF#
Gets/Sets the frequency that the continuous strobe operates at.
Input Options (#): 6 – 20000 = Frequency in Hz, ? = Query
Return string:
&rfX
Return value (
X
): 6 – 20000 = Frequency in Hz
3.3.4
Duty Cycle
3.3.4.1
Legacy
&RD#
Gets/Sets the time that the LED is active during the strobe event.
Input Options (#): 0 – 1000 = Duty Cycle, ? = Query
Return string:
&rdX
Return value (
X
): 0 – 1000 = Duty Cycle
3.3.4.2
CV-LS
&RD#,#
Gets/Sets the time that the LED channel is active during the strobe event.
First Input Option (#): 1 – 4 = Channel
Second Input Option (#): 0 – 1000 = Duty Cycle, ? = Query
Return string:
&rdX,Y
Return value (
X
): 1 – 4 = Channel
Return value (
Y
): 0 – 1000 = Duty Cycle