6
2.2 Delivery Includes
1 Tension meter
1 Certificate of Compliance with the order 2.1 under EN 10204
1 Operating Instruction
1 Carrying case
2.3 Unpacking
Unpack the tension meter and inspect it for any shipping damage. Notices of defect must
be announced immediately, in writing, at the latest within 7 days on receipt of the goods.
3 Initial Setup and Operating Procedure
3.1 Notes Before Starting Measurement
Have you read and understood the Operating Instructions, in particular
Chapter 1 “Basic Safety Notices” ?
You are not permitted to operate the tension meter before doing so.
Before working with the instrument you must put on your personal protective
clothing, if necessary. For example, eye protectors, gloves, etc.
To avoid damage, do not move the center roller by hand.
Tensions that exceed the tension range of the instrument by more than 100%
may cause permanent damage to the measuring spring and must be avoided
under any circumstances.
Do not turn the outer ring of the movement, the measured values will oth-
erwise be wrong. If the outer ring has been turned, check the calibration as
described in Chapter 3.3 and correct the position of the outer ring.
The ID plate with the serial number as well as the calibration label (optional) are
provided on the rear side of the instrument, the SCHMIDT Quality Seal are pro-
vided on the front side.
fig. 3.1
Calibration label
(option)
SCHMIDT Quality Seal
ID plate with serial number