11
3.4.2.1 Delay Mode
rpm
: Light will flash in correspondence with
input signal; the input signal will be
calculated into rpm and displayed.
deg
: One cycle of input signal is 360°.
A delayed angle will be displayed from 0
up to 359°. (The delayed angle can be
changed by turning the setter as
previously described).
msec
: The above delayed angle will
be displayed in msec.
If the input signal frequency exceeds upper or lower limits, the alarm dashes
(
)
will be displayed and the strobe will stop flashing (see chapters
3.4.2.3 and 3.4.2.4).
If the Auto Power Off timer is used:
Once the internal timer has expired, the strobe will stop flashing and the display
will flash rapidly. To restart the strobe, turn power switch off, then on, and the
cycle will repeat. (see chapter 3.5)
3.4.2.2 RPM Display Mode
If the input signal exceeds 585 Hz, the upper dashes on the digital display
will be flashing:
upper dashes
If the input signal is lower than 0.65 Hz, the lower dashes on the digital
display
will be flashing:
lower dashes
3.4.2.3 Deg/msec Display Mode
If the input signal exceeds 167 Hz, the upper dashes on the digital display
will be flashing:
If the input signal is lower than 0.67 Hz, the lower dashes on the digital
display
will be flashing:
upper dashes
msec
deg
deg
lower dashes
msec