Manual cleaning
Do not place the contra-angle in liquid disinfectant or in an ultrasonic bath
1. Rinse contra-angle under running tap
water.
(<35°C / 95°F).
Rinse and brush off
all internal and external surfaces.
Move moving parts back and forth several
times.
2. Remove any liquid residues using
oil-free,
filtered
compressed air.
Manual disinfection
We recommend wiping down the contra-angle with disinfectant
Evidence of the medical device’s basic suitability for
effective manual disinfection was provided by an
independent test laboratory using the »mikrozid®
AF wipes« disinfectant (Schülke & Mayr GmbH,
Norderstedt, Germany).
If this product is not available, »CaviWipes®« dis-
infectant (Metrex LLC,
Orange, CA 92867 USA
)
is an alternative.