Pilot²
Doc. No. 610 2221 v07
- 31 -
(*applicable with the ultrasonic extension module)
6. Ultrasonic function*
6.2 Setting of the ultrasonic power output
Select the power by pushing the bar graph
6.4 Ultrasonic instrument selection
Press the name of the tool to enter the
tool-library
6.4 Setting of the run time
Select the run time directly at the clock
With the motor button you will return
to the last used file system
6.1.
Operating instructions*
When inserting the instruments, make sure that only instruments with a suitable type of
thread are used.
First tighten the selected instrument by hand and then with the wrench. When using fixed
wrenches, take care not to overtighten the instruments. The handpiece and the instrument
may be damaged. Only change the instruments when the ultrasonic handpiece is switched
off. Do not twist the handpiece’s cable connection. Do not use instruments that have been
deformed or show wear. Pay attention to the product life cycle of the instruments. Work
without exerting excessive pressure on the instruments. Instrument breakage may lead to
injuries. Dispose of used instruments in an appropriate manner. Ensure that a contamination
or an infection is prevented.
If possible, cool the instrument with external cooling to prevent it and the treatment site from
heating up. Always use sterile water when working in the root canal.
Please note that freely vibrating, thin instruments may also break without an external load,
which is solely due to the ultrasonic vibration. Flying splinters may lead to injuries. The device
may only be operated when the ultrasonic handpiece is connected to the handpiece cable.
Never touch the handpiece cable’s contacts! Select only appropriate program parameters for
the instrument. You must no longer used the device if malfunctions or disturbances occur.
Contact the manufacturer immediately.
Ensure that the ultrasonic tips are used according to your field of application. Do not use any
other ultrasonic handpieces or cables.
During the ultrasonic application, the patient must no longer be connected to the apex cable!
Place the handpieces back into the holder after use.
The handpiece connection of the ultrasonic handpiece to the cable and the internal contacts
must be absolutely dry. Pay attention to damage to the insulation and the handpiece. Only
use proper functioning handpieces and ultrasonic cables. Damaged insulation may lead to
electric shocks.