![Schlumbohm 110 0607 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/schlumbohm/110-0607/110-0607_user-manual_1212770019.webp)
Pilot²
Doc. No. 610 2221 v07
- 19 -
(*applicable with the ultrasonic extension module)
2.6.
Charging, switching-on, standby mode, switching-off
Make sure to fully charge the device before first use. (The device can only be charged or
switched on with the inserted microSD card.)
When charging, please ensure that the device has not been heated by sunlight. Charging is
interrupted at a device temperature above 40°C.
To charge, plug the power supply unit into the socket (the green LED in the power supply
unit must light up). The device plug of the power supply unit is plugged into the black socket
(1c) on the rear of the device. The device is switched on automatically by connecting the
power supply unit, the blue LED on the front of the device flashes during charging.
During charging, the display illumination can be switched off with the On/Off switch on the
rear of the device, charging continues. When the battery is fully charged, the blue LED lights
up continuously. The power supply can be disconnected.
The respective battery status is displayed at the bottom edge of the screen.
If the charge drops to 10% of the capacity, a warning message appears. In this case the
battery must be charged immediately. If not charged, the device will switch off to avoid a total
discharge of and damage to the battery.
Charge the battery regularly.
If the device is not used for a prolonged period of time, the device automatically switches to
sleep mode and the display illumination switches off. The sleep mode is indicated by slow
flashing of the blue LED in the display. By briefly pressing the foot switch or the touch display,
the device switches on again. The last menu used is displayed again.
After a long waiting period, the device will switch off completely. This "Auto off" time can be
set in the setup menu.
To avoid unnecessary power consumption in standby mode, the mains plug should be
removed from the plug socket when the EndoPilot² is not in use for a longer period of time.
In case of malfunctions, you can completely switch off the device by removing the microSD
card. The microSD card must be removed when the device is shipped.
2.7.
Preparation of the root canal - motor and contra-angle
The EndoPilot² contra-angle (5) is attached to the motor (6). Only use contra-angles with a
1:1 ratio. The integrated apex length determination during preparation
(see chapter 4.5.3) only works in conjunction with the original EndoPilot² contra-angle.
If the contra-angle was changed or sterilized, a calibration must be performed under the
Calibration
(motor menu) menu item. The calibration compensates the friction in the
contra-angle. Contra-angles may only be changed when the motor is at a standstill.
Operating instructions:
Before operation, check that the motor is firmly locked in place in the contra-angle.