DEFIGARD/PHYSIOGARD Touch 7
5.5.2
Voice messages in AED Mode
The following instructions will be spoken by the device:
The sequencing of AED instructions might be very fast and could cause
confusion to the user because:
The “Anteriority
Analyse” feature pre-analyses the heart rhythm before the
actual analysis. This feature can substantially reduce the duration of the
analysis.
Spoken instructions
Display
Note
Plug and apply electrodes
Illustration for electrode
connection
Technical alarm:
Electrodes not yet applied. The message
disappears as soon as the electrodes are
correctly applied and the resistance is between
25 to 250 Ohm.
Do not touch the patient. Analysing
DO NOT TOUCH THE
PATIENT ANALYSIS
Movement detected - Analysis
cancelled, resume CPR
Movement detected -
Analysis cancelled, resume
CPR
Technical alarm: Patient was moved during
analysis and device could not run analysis.
Device recommends a shock
Shock advised
Stand clear of patient; press orange
button
Stand clear of patient, press
orange button
TO SHOCK
Device does not recommend a shock
No shock advised
No shock advised.
Immediately resume CPR: 30 chest
compressions, then 2 rescue breaths –
continue until patient is breathing
normally.
30
a
CHEST
COMPRESSIONS
THEN 2 RESCUE
BREATHS
a.When paediatric electrodes are used, CPR is carried out in the ratio of 15:2 if 2 rescuers are on the spot, otherwise 30:2.
A “continuous compressions" option is also available (i.e. no rescue breaths)
PLUG AND