Initial Operation
3.1
External DC supply and Battery Operation
Page 26
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
3 Initial Operation
3.1 External DC supply and Battery Operation
3.1.1
External DC Supply Operation
Please read the safety notes in section
operation.
Danger of explosion! The device is not designed for use in areas where an
explosion hazard may occur. Also, it is not permitted to operate the defibrillator
in an oxygen-enriched environment or in the presence of flammable substances
(gas) or anaesthetics. Oxygenation in the vicinity of the defibrillation electrodes
must be strictly avoided.
Danger of electrical shock. The
DEFIGARD
®
Touch 7
is a high-voltage therapy
device. Improper use of the device can endanger life. Always follow the
instructions given in this user guide.
The user must make sure that there are no conductive connections between the
patient and other persons during ECG analysis and defibrillation.
Avoid defibrillation in very moist or wet surroundings.
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
1. Put the device into the docking station. Insert a fully charged battery. Check the
LED
2
is on if placed on the docking station.
2. Press the
on/off
button.
3. Touch the battery icon (3) to display further battery charging information.
4. Check battery charging LED
1
according to
3.1.2 Battery Operation page 27.
Fig. 3.1
Status LED supply
2
3
1