Maintenance
10
User guide
Functional test
10.2
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
10.2.4
Auto Test
The Auto Test can be executed any time and checks the most important function of
the device.
1. Switch on the device
2. Select
Menu > Post-Intervention > Auto Test
3. Start the auto test.
4. If message “Press shock button” appears, press the button
to continue and
finalizing the test.
Previous tests are listed in the sub-menu “Review/send previous tests”.
10.2.5
Functional test - measured values
Heart rate
Perform the functional test according to section
page
and verify the heart
rate with the measured pulse rate of the SpO2.
SpO2
This test is performed on a volunteer (finger measurement; section
page
NIBP
This test is performed on a volunteer (arm measurement; section
)
Manometer Test
See technical manual 0-48-0245_NT_DGTouch7_ANG
IBP
This test only includes the connection between the sensor and the Touch 7, and the
Zeroing function.
1. Connect the sensor to the Touch 7 according to section
.
2. Perform a “zeroing according to section
page
.
3. Zeroing is shown on the display.
CO2 Mainstream
Perform the functional test according to section
page
CO2 Sidestream
Perform the functional test according to section
page
10.2.6
Alarm tests
Alarm volume
Check during the following tests that the alarm sound is higher than 65 dB.
Heart rate
1. Start the ECG monitoring (see section
, page
2. Set the alarms with the narrow quick set function (see section
, page
3. When the measured value exceeds the alarm thresholds, an alarm is issued.
4. Reset the alarm limits to their original values.
SpO2
See section
50.
NIBP
1. Start the NIBP monitoring (see section
2. Set the NIBP alarm limits below/above the measured values and take a new
measurement.
3. When the measured value exceeds the alarm thresholds, an alarm is issued.
4. Reset the alarm limits to their original values.
CO2
1. Start the CO2 monitoring according to section
page
page
2. Set the alarms with the narrow quick set function (see section
, page
3. When the measured value exceeds the alarm thresholds, an alarm is issued.