Page 105
Pacemaker
6
User guide
Start-up of the Pacemaker
6.4
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
6.4.4
Demand Mode
1. Attach the pacer pads (see
2. Display pacemaker and select operational mode Demand.
3. Select
(1)
Fre-
to set the impulse frequency.
4.
Starting the pacemaker!
Press the
Pacer OFF/ON (2)
to activate the pacemaker.
5. Press the
(3) C- mA
to set the impulse current until the heart reacts to
the stimulation.
6. The pacemaker can be interrupted and restarted by selecting the
Pacer OFF/ON
.
7. Finish the therapy as described in section
7 Finishing the Therapy page 106
.
6.4.5
Switching from pacemaker to defibrillation
To determine when a pacemaker pulse is required, the patient ECG must be
monitored with a 4- or 10-lead ECG cable.
Shock hazard!
Pacing is started immediately when the pacemaker is switched ON and the
current is set.
Never touch the pads or the patient's body near the pads while the pacemaker
is in use.
1
3
2
Frequency
Current
Fixed
Demand
1. Press the AED key.
2. Use the same key to confirm stopping the pacemaker and switching to defibrilla-
tion mode.