![Schiller CARDIOVIT CS-104 Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/schiller/cardiovit-cs-104/cardiovit-cs-104_user-manual_1212080076.webp)
7
Exercise ECG Recording
7.5
During the Test
Page 76
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
7.5.1
Test Control Buttons
The test control button on the bottom of the screen are as follows:
Hold / Pause (treadmill)
Hold current stage. The button changes to
Resume
and the step counter is stopped
(but total work counter continues) until the stage is resumed. During this time the load
on the bike remains / the treadmill speed continues, until resume is again clicked.
Next
Go to next stage.
Init Recovery / Recovery
(depends on
setting)
Initiates the recovery phase and the button changes to End. The recovery stage is
held for as long as defined in the protocol or until End button is pressed.
Search
Returns to the search and discontinues the current ECG acquisition. You are
prompted to confirm:
Acquisition ongoing
An acquisition is currently ongoing. Do you want to discard
the current recording?
Discard
Cancel
7.5.2
Information and Settings Panel
Heart Rate
Averaged over 4 beats.
• The percentage of current heart to target heart
rate is displayed after the heart rate (the formula
used to define the target heart rate is defined in
system settings (
.
120
48% of 170
82
• If the target is exceeded, the Heart rate indica-
tion changes from green to red and flashes to in-
dicate that the target is exceeded.
Blood Pressure
Last entered/recorded blood pressure measurement (systolic/diastolic). The value is
displayed until a new value is entered /measurement taken, or until a new exercise
stage is entered. The previous BP is given beside the current value.
mmHg x bmp
The (last) BP measurement taken in the stage, multiplied by the heart rate.
Lactate
Select to enter the respective measurement and description.
SpO
2
The SpO
2
measurement received from a connected device /entered manually. (
VEBR
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...