Recording an ECG
4
User Guide
Placing the Electrodes
4.1
A
rt.
no.
:
2.
511
335
Rev.
a
CARDIOVIT CS-104
4.1.12
Frank Leads X, Y, Z
The orthogonal lead configuration is based on the theory of the heart as centre of a
three-dimensional system of coordinates:
lateral axis X
longitudinal axis Y
sagittal axis Z
With the patient lying down, attach the electrodes on a level with the fourth intercostal
space. With a patient in a seated position, attach the electrodes in the fifth intercostal
space.
Place all other electrodes in the normal position
s.
C1 white / red
V1 brown / red
I (-X) -
right midaxillary line
C2 white / yellow
V2 brown / yellow
E (-Z) -
front midline
C3 white / green
V3 brown / green
C (+Y) -
between E (-Z) and A (+X)
C4 white / brown
V4 brown / blue
A (+X) -
left midaxillary line
C5 white / black
V5 brown / orange
M (+Z) -
back midline (on the back)
C6 white / violet
V6 brown / violet
H (-Y)
- neck (on the back)
L yellow
LA black
Standard position - left arm
R red
RA white
Standard position - right arm
F green
LL red
Standard position - left foot
N black
RL green
Standard position - right foot
IEC Label
AHA Label
Electrode Placement
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...