11
System Settings
11.5 Import / Export
Page 126
Art
.
no.
:
2.
51
1335
Rev.
a
CARDIOVIT CS-104
Filter (Resting, Rhythm, and Exer-
cise ECG)
• Off - Recording is exported with no filtering.
• LP 25, 40, 150 - recording is exported with l
ow pass filter
set to the defined cut-
off frequency.
• RNSF (exercise only) - recording is exported with RNSF (Robust noise suppres-
sion filter
1
).
11.5.3
GDT /SEMA2
Character set
IBM (standard) CP437, ISO8859-1(ANSI) CP 1252, ASCII, etc. This is set to the
character set defined by the EMR system
EMR Device
The ID of the remote system with which the CARDIOVIT CS-104 will communicate.
This is defined by the remote system.
Device ID
The device ID identifies CARDIOVIT CS-104 in the system to enable communication
between CARDIOVIT CS-104 and the remote system. The ID must be unique in the
system and can be any combination of up to four characters.
Because all information is assigned using these IDs, it is essential that all IDs be
unique, especially in the case of multiple devices (for example 2 ECG devices from
the same manufacturer).
File extension use
Select
auto increment
or
static
. When auto increment is selected, the file extension
will increase by one every export, that is 0001, 0002, etc.
Export Measurements
This setting is how and where to export the measurement table and is set according
to GDT system requirements. The options are as follows:
– After Interpretation - append the measurement table after the interpretation
statement
– No Export - do not export the measurement table
– Pre-formatted result table
– Re-occurring test parameter fields
Details of these options are given in the SCHILLER Server communication guide.
1. The RNSF filter is a dual purpose filter specifically used for exercise testing. It is designed
to help stabilise the baseline during exercise tests and to help reduce the physical and mus-
cle artefacts that can be present during exercise testing. The RNSF addresses artefact fre-
quencies both outside and within the ECG frequency range.
Содержание CARDIOVIT CS-104
Страница 1: ...User Guide CARDIOVIT CS 104 Art no 2 511335 Rev a 2 511335 ...
Страница 12: ...Page 12 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 170: ...19 Annex Installation 19 5 Connection Overview Page 170 Art no 2 511335 Rev a CARDIOVIT CS 104 ...
Страница 172: ...Page 172 Art no 2 511335 Rev a 20 Index CARDIOVIT CS 104 ...