5.3 Functional checks and tests
5.3.1
External sight control
6-monthly
Check the following :
• Mechanical condition of the device:
– no cracks or chips in the casing.
– mains, patient and all other cable assemblies are in good condition with no
crushing, chafing or cuts, etc.). All plugs and sockets are straight and in good
condition.
• no soiling which could hamper the safety of the device.
12-monthly
Check the following:
• Mechanical condition of the device as detailed above.
• Voltage selector is set correctly.
• Correct fuse rating according table (
• The following safety labels are on the device and are readable:
– Back Panel, type designation and fuse rating label.
– Side Panel (patient connector), CF label and ‘attention‘ symbol.
Required equipment
• Safety Tester IEC/EN 60601-1
• Calibrated ECG Patient Simulator (e.g Müller & Sebastiani MS410 ECG Simulator)
IMPORTANT!
The measurement devices listed above are subject to the instructions according to
ISO 9000 in regards to Test Equipment Control.