RTC
®
4 PC Interface Board
Rev. 1.3 e
10 Commands And Functions
77
innovators for industry
Ctrl Command control_command
Function
sends a control command to intelli
SCAN
®
scan systems
Parameter
head
= 1:
Head A
= 2:
Head B
axis
= 1:
X axis (STATUS channel, galvanometer scanner 2)
= 2:
Y axis (STATUS1 channel, galvanometer scanner 1)
= 3:
Z axis
data
Command code as unsigned 16-bit value for selecting the data signal to be
returned to the selected axis
(signed 16-bit values are generally returned)
Code
Returned Data Signals
0500
H
Statusword (Status signal as per XY2-100-Protocol) [0000
H
... FFFF
H
]
Bit #15 (MSB), Bit #7
= 1:
Internal voltages normal
Bit #14, Bit #6
= 1:
Galvo temperature within normal range
Bit #12, 11, 4, 3
= 1:
X- and Y-axis position error within normal range
Bit #13, 10, 8, 5, 2, 0
= 1
Bit #9, 1
= 0
0501
H
Actual (angular) position / bit [–32768 ... 32767]
0502
H
Set (angular) position / bit [–32768 ... 32767]
0503
H
Position error (= set position - actual position) / bit [–32768 ... 32767]
0504
H
Actual current (output stage current) / mA [–32768 ... 32767]
0505
H
Relative galvo control / ‰ [–1000 ... 1000]
0506
H
Actual (angular) velocity / (bit/ms) [–32768 ... 32767]
0514
H
Galvanometer scanner temperature / 10
-1
°C [–32768 ... 32767]
0515
H
Servo board temperature / 10
-1
°C [–32768 ... 32767]
0516
H
AGC voltage (PD supply voltage) / 10
-2
V [–32768 ... 32767]
0517
H
DSP core supply voltage (1.8 V) / 10
-2
V [–32768 ... 32767]
0518
H
DSP IO voltage (3.3 V) / 10
-2
V [–32768 ... 32767]
0519
H
Analog section voltage (9 V) / 10
-2
V [–32768 ... 32767]
051A
H
AD converter supply voltage (5 V) / 10
-2
V [–32768 ... 32767]
051B
H
AGC current (PD supply current) / mA [–32768 ... 32767]
051D
H
Relative heating output of the corresponding galvanometer
scanner heater /
‰
[0 ... 1000]
051E
H
Serial number (lower 16 bits) [0 ... 65535]
051F
H
Serial number (higher 16 bits) [0 ... 65535]
0520
H
Article number (lower 16 bits) [0 ... 65535]
0521
H
Article number (higher 16 bits) [0 ... 65535]
0522
H
Firmware version number [0 ... 65535]
0523
H
Calibration / mrad [0 ... 65535]
0524
H
Aperture / mm [0 ... 65535]
0525
H
Wavelength / nm [0 ... 65535]