36
MPS4264
Section 5: Software
Formatted ASCII Data Output
“Table 5-3: Formatted ASCII Data Output” shows an example of ASCII data output formatted for a VT100 Terminal
(SET FORMAT T F):
Table 5-3: Formatted ASCII Data Output
Frame= 1
T1= 35.1 C T2= 34.7 C T3= 35.1 C T4= 34.9 C
T5= 35.1 C T6= 34.8 C T7= 34.7 C T8= 34.9 C
01= 0.0028 02= -0.0016 03= -0.0011 04= -0.0014 05= 0.0020 06= -0.0003
07= -0.0003 08= -0.0014 09= 0.0017 10= 0.0011 11= -0.0002 12= -0.0009
13= 0.0032 14= 0.0015 15= 0.0028 16= -0.0005 17= 0.0018 18= 0.0022
19= -0.0006 20= -0.0019 21= -0.0000 22= -0.0007 23= -0.0018 24= -0.0022
25= 0.0029 26= 0.0001 27= 0.0000 28= -0.0006 29= -0.0000 30= 0.0010
31= 0.0004 32= -0.0007 33= -0.0005 34= -0.0018 35= 0.0008 36= 0.0011
37= 0.0003 38= -0.0001 39= -0.0001 40= 0.0029 41= -0.0005 42= -0.0016
43= -0.0003 44= 0.0019 45= 0.0003 46= -0.0002 47= 0.0001 48= 0.0021
49= -0.0002 50= 0.0003 51= 0.0004 52= 0.0016 53= 0.0014 54= 0.0002
55= 0.0009 56= 0.0007 57= -0.0006 58= -0.0010 59= -0.0009 60= 0.0024
61= -0.0008 62= -0.0006 63= 0.0002 64= 0.0026
Notes:
Configuration selected with by setting the “FORMAT” variable to “F”.
This format should only be selected for displaying data on a screen.
The 8 temperatures are shown as T1 through T8.
The 64 pressures are shown as 01 through 64
This format uses VT-100 control codes to set cursor positions for column alignment.
ASCII CSV Data Output
“Table 5-4: CSV Data Output” shows an example of Comma Separated Variable data formatted (SET FORMAT T C):
Table 5-4: CSV Data Output
Frame,Seconds,Tx1,Tx2,Tx3,Tx4,Tx5,Tx6,Tx7,Tx8,Px1,Px2,Px3,Px64
1,0.100,40.38,39.88,40.38,40.13,40.31,40.00,39.88,40.13,0.0030,-0.0011,-0.0010,-0.0025
2,0.200,41.50,41.00,41.50,41.25,41.44,41.13,41.00,41.25,0.0022,-0.0013,-0.0014,-0.0024
3,0.300,41.50,41.00,41.50,41.19,41.44,41.13,41.00,41.25,0.0017,-0.0017,-0.0006,-0.0030
4,0.400,41.50,41.00,41.50,41.19,41.44,41.13,41.00,41.25,0.0013,-0.0018,-0.0005,-0.0026
5,0.500,41.50,41.00,41.50,41.19,41.44,41.13,41.00,41.25,0.0021,-0.0013,-0.0005,-0.0028
6,0.600,41.50,41.00,41.50,41.19,41.44,41.13,41.06,41.25,0.0025,-0.0015,-0.0009,-0.0031
7,0.700,41.50,41.00,41.50,41.19,41.44,41.13,41.06,41.25,0.0027,-0.0011,-0.0015,-0.0026
8,0.800,41.56,41.00,41.50,41.19,41.44,41.13,41.06,41.25,0.0026,-0.0004,-0.0015,-0.0025
Notes:
Configuration selected with by setting the “FORMAT” variable to “C”.
The output shown omits channels 4 through 63 for simplicity
Each column is separated by a comma and each row is delimited by a CR LF
This format is easily imported to Excel or Open Office spread sheets.
Содержание MPS4264
Страница 1: ...MPS4264 Miniature Pressure Scanner Hardware and Software Manual Software Version 2 07...
Страница 2: ......
Страница 3: ......
Страница 6: ...INTENTIONALLY LEFT BLANK...
Страница 12: ...6 MPS4264 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Страница 18: ...12 MPS4264 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Страница 32: ...26 MPS4264 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Страница 38: ...32 MPS4264 Section 5 Software INTENTIONALLY LEFT BLANK...
Страница 81: ...75 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Страница 86: ...80 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Страница 87: ...81 MPS4264 Appendix INTENTIONALLY LEFT BLANK...
Страница 98: ...92 MPS4264 INTENTIONALLY LEFT BLANK...