60
MPS4232
Section 5: Software
SYNTAX
CALVAL
[F]
Where:
F = Initiates a single unit field calibration, followed by a field validation with no oven connec
-
tion. MPS connection to a calibrator required.
When the optional parameter is not present, the unit performs a full calibration/validation.
RETURNS
Calibration and validation information and a prompt when completed.
EXAMPLE
CALVAL
[:::data:::]
>
NOTES
A Scanivalve SPC4000 or SPC4050 is required for automated calibrations/validations.
An Ethernet controllable thermal chamber is required for full calibration/validations.
Multiple Unit Calibration & Validation
COMMAND
MCALVAL
TYPE
Function
DESCRIPTION
When the master receives the MCALVAL command, it starts the multiple device full or field cali
-
bration and validation sequence (continuous), as defined by the variables in LIST C and/or LIST O.
A validation report is generated and stored in each MPS’s file system.
SYNTAX
MCALVAL
[F]
F = Initiates a multiple unit single temperature calibration with no oven connection. MPS
connection to a calibrator required.
When the optional parameter is not present, the unit performs a full multi-unit calibration and
validation.
RETURNS
Calibration and validation information and a prompt when completed.
EXAMPLE
MCALVAL
[:::data:::]
>
NOTES
A Scanivalve SPC4000 or SPC4050 is required for automated calibrations/validations.
An Ethernet controllable thermal chamber is required for full calibrations/validations.
Validation
COMMAND
VAL
TYPE
Function
DESCRIPTION
Starts a full or field validation, as defined by the variables in LIST C and/or LIST O. A validation
report is generated and stored in the MPS’s file system.
SYNTAX
VAL
[<F or FM>]
Where:
F = Initiates a single temperature field validation, with no oven connection. MPS connection to
a calibrator required.
FM = Initiates a single temperature manual field validation with no oven or calibrator connec
-
tion.
When the optional parameter is not present, the unit performs a full validation.
*Coming in a future firmware update.
RETURNS
Validation Data
EXAMPLE
VAL
[:::data:::]
>
Содержание MPS4232
Страница 1: ...MPS4232 Miniature Pressure Scanner Hardware and Software Manual Software Version 1 00...
Страница 2: ......
Страница 3: ......
Страница 6: ...INTENTIONALLY LEFT BLANK...
Страница 12: ...6 MPS4232 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Страница 15: ...9 MPS4232 Section 1 Introduction Section 1 Specifications MPS4232 INTENTIONALLY LEFT BLANK...
Страница 18: ...12 MPS4232 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Страница 33: ...27 MPS4232 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Страница 37: ...31 MPS4232 Section 5 Software INTENTIONALLY LEFT BLANK...
Страница 51: ...45 MPS4232 Section 5 Software...
Страница 81: ...75 MPS4232 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Страница 90: ...84 MPS4232 Appendix INTENTIONALLY LEFT BLANK...
Страница 94: ...88 MPS4232 Appendix D Software Revision Log Version 1 00 Initial Release...
Страница 95: ...MPS4232 89 Appendix INTENTIONALLY LEFT BLANK...