AccuPro-5000
™
MEASUREMENT SYSTEM OPERATOR MANUAL
Issued by Department:
Engineering
Doc. Number:
5000-00-00-002
Revision:
02
Effective Date:
10/05/2021
File: 5000-00-00-002-r2_
AccuPro_5000_Operator_Manual
Copyright© 2021 Scaletron Industries, Ltd
All Rights Reserved
Page
20
of
63
Figure 4: Analog Input Board
4.4.1
Features:
•
Leak Detection Interface
: A 3-pin port that supplies a +5 VDC power source to an optional,
external fluid sensor located on a scale base. An LED indicator reflects the real-time status of the
fluid sensor; however, any assertion of the leak detection circuit will be latched as an Alarm
condition which requires a user response in order to reset.
•
Load-Cell Interface
: A 4-pin port that supplies an excitation voltage to one or more load-cell
bridges and receives back their summed output signal.
•
Flash Memory
: Non-volatile memory space used to store channel-specific calibration and
configuration data for both the AIB and ROB.
•
Bridge Transducer
: A 24-bit Sigma Delta Analog-to-Digital Converter (ADC) that filters,
amplifies, and converts a low-voltage, analog load-cell signal into a digital value.
•
Excitation Voltage Regulator
: Configured for +7.5 VDC @ 184 mA (maximum.) Offers ultra-
low noise performance with a “power good” status output signal.
4.5
Analog Output Board (4-20 mA Output)
With reference to Figure 6, the AOB-mA is an optional board that supports an industry standard, 4 to
20 mA current loop output. Its instantaneous current output is linearly proportional to the
measurement value
’s
relative position between two user-defined limit values. External wire
connections are made to the 2-position terminal block according to
Leak
Detection
Sensor
Interface
Load Cell
Interface
Flash
Memory
Bridge
Transducer
Excitation
Voltage
Regulator