POWEO® 150 

PATIENT LIFTER 

MAXIMUM LOAD : 150 KG

 

User Manual 

Содержание POWEO 150

Страница 1: ...POWEO 150 PATIENT LIFTER MAXIMUM LOAD 150 KG User Manual...

Страница 2: ...call SCALEO Medical which will offer you the hoist that best suits your needs The POWEO 150 is not intended to be equipped with a rigid stretcher the installation of such a device instead of a flexib...

Страница 3: ...actuator with manual emergency down function 6 removable battery with emergency stop button 7 remote control 4 or 6 buttons depending on the equipment 8 User manual not present in the picture below 9...

Страница 4: ...Medical for further information Using the POWEO 150 patient lifter require a good understanding of transfer systems assembly and disassembly procedures SCALEO Medical tests and charges all batteries...

Страница 5: ...aise the column and then insert the axis to fix the column and secure it by blocking the screw tightly using the special wrenches If needed ask the second operator to help you by moving slightly the c...

Страница 6: ...p them for the future they may be useful for later servicing of the POWEO This operation is very important for the security of the patient and caregivers Without the security pin the column may fall a...

Страница 7: ...soon as the battery is charged Whenever possible keep the carton bow and packing accessories for later transportations 2 Instructions for storing and transportation For foldable version follow the ass...

Страница 8: ...features raise and down the arm leg opening emergency stop In case of malfunction or doubt about the proper operation do not use the POWEO inform your technical service your distributor or contact us...

Страница 9: ...he button to obtain the wanted height In case of a complete failure of electronics a MANUAL emergency lowering is available Activate the RED lever trigger on the actuator by pulling it upward and main...

Страница 10: ...d value the battery will stop functioning preventing the user from overheat and fire hazards Therefore the use of SCALEO Medical batteries and chargers is perfectly secure In order to recharge the bat...

Страница 11: ...mergency stop button Keep your battery charged in cool dry well ventilated Check periodically every three months the state of charge of the batteries even disconnected batteries may self discharge Com...

Страница 12: ...to indicate a possible malfunction AND the periodic inspection state Therefore it can display multiple consecutive states Example in case of default AND periodical inspection date exceeded the LED di...

Страница 13: ...commended to use others slings from others manufacturers than the ones recommended for SCALEO Medical patient lifters SCALEO Medical refuses any liability in the case of an incident while using slings...

Страница 14: ...nt of the room With or without head support the PRAXIS Healthcare slings are in a U shape Slings should never be bleached The washing conditions and temperatures are showed on the label Fold the sling...

Страница 15: ...ops one through the other NB If the patient s bed can be lowered it should be lowered as much as possible before the lifting starts Pull the leg loops forward and under the thighs as shown Position th...

Страница 16: ...he back of the sling Raise the patient until his buttocks are just above the mattress Lift the user s legs and turn him so he faces the lifting column Once away from the bed and if possible lower the...

Страница 17: ...ulders Place the folded sling behind his back and push it down until it touches the seat of the chair The leg sections of the sling are pulled forward and beneath the user s thighs by the loops The lo...

Страница 18: ...knees pull the handle at the back of the sling Slightly tilt the chair backwards Lower the patient into the chair unhook the sling and remove the lifter Removing the sling Gently pull the leg section...

Страница 19: ...t s back with the grey side away from him 2 Fasten the belt firmly around the patient s waist 3 Place the leg section over and down between the thighs Pull outwards from the inner part of the thighs A...

Страница 20: ...gly advised NOT to use straps from other manufacturers If that were the case SCALEO Medical disclaims any responsibility for any incident or accident Slings dimensions The dimensions given are approxi...

Страница 21: ...SLINGS REFERENCES...

Страница 22: ...uction of hazardous substances A EU declaration of conformity can be provided on demand The Poweo 150 and Poweo 150 Low lifters are entirely foldable and have a locking system to keep all parts secure...

Страница 23: ......

Страница 24: ...th outside 68 9 cm Outspread legs width outside 125 5 cm Overall length 130 2 cm Maximum height while unfolded 188 5 cm Chassis height 10 3 cm Minimum space under unit 3 3 cm Hoisting range 155 cm Max...

Страница 25: ...POWEO 150 LOW BASE TECHNICAL SPECIFICATIONS...

Страница 26: ...th outside 68 9 cm Outspread legs width outside 125 5 cm Overall length 130 2 cm Maximum height while unfolded 188 5 cm Chassis height 6 9 cm Minimum space under unit 2 1 cm Hoisting range 155 cm Maxi...

Страница 27: ...cm Overall length 130 2 cm Maximum height while unfolded 194 9 cm Chassis height 10 3 cm Minimum space under unit 3 3 cm Hoisting range 155 cm Maximum reach at 600 mm 69 6 cm Maximum reach from base...

Страница 28: ...z or 60Hz The only adaptation needed is the AC cord Before ordering please inform our services of the place of use SCALEO Medical will provide the good model of the AC cord Input AC voltage 100 240VAC...

Страница 29: ...t is used In terms of the standard NF EN 60601 1 2 2007 essential performance of the POWEO 150 is to never perform unwanted movements Directives and manufacturer s declaration electromagnetic emission...

Страница 30: ......

Страница 31: ......

Страница 32: ...or poison control centre Recommended frequency for the cleaning and disinfection The multiple use devices should be thoroughly cleaned and or disinfected at least once a month if they are used by one...

Страница 33: ...places raised above ground and protected from dust moisture heat sunlight and risk of deterioration of packagings tears staples Product should be avoided for cleaning and disinfecting peracetic acid a...

Страница 34: ...the equipment must be done regularly every year so as to prevent malfunctions that could harm the patients or users according to annexe B of the EN ISO 10535 standard PERIODIC CONTROL PREVENTIVE MAIN...

Страница 35: ...cal inspection date has expired The main fuse it out of order real time clock out of order or not right The main actuator is 97 worn and must be changed Storage capabilities Date of first use first po...

Страница 36: ...alarm is on Using functions without explanation Use of the lifter or its accessories by children Using the patient lifter with a higher than 150 kg load Outdoor use Use with or without a patient on a...

Страница 37: ...option This warranty does not include any shipping charges incurred in replacement part installation or repair of such product Our sole obligation and your exclusive remedy under this warranty shall b...

Страница 38: ...only Do not open or unmount Contains potential wastes of electrical or electronic equipment Read user manual for instruction Direct current power source CE mark the POWEO is compliant to European regu...

Страница 39: ...1800 337 711 03 9314 6300 Sales Enquiries sales warequip com au Contact Information Service and Spare Parts service warequip com au Customer Service customerservice warequip com au www warequip com au...