Data recorded space blocks indicate condition results (
P
: Pass,
W
: Warning,
F
: Fail,
G
:
Gravity Only,
O
: No Assigned Reference). A blank space indicates no data is stored in that
location.
When you select a string, the reference data for the string appears at the bottom.
Move the cursor to select the block where you want to begin storing readings and press
the F1 “Start” key to enter the resistance measuring mode.
7.4.
String Set-Up
String list is the identification of a battery system that records are going to be saved in.
[String Addition on the SBS-6500]
①
Press
key to enter data display screen.
②
Press
key to select “Record” and press
to enter.
③
Use
key to move to vacant string number.
④
Press the
key to open the new string entry screen
⑤
The cursor highlights the string name (default value is Unit ###).
⑥
Press the
EDIT key to open the keyboard. Then enter a unique string name to
identify the battery system and location where the readings will be taken .
⑦
Use
key to cursor move to REF DB..
⑧
Use
keys or rotary selector to select reference data from DB that you want
to use as the alarm set points for that string.