SeapHOx

 

Satlantic SeaFET

 pH Sensor integrated with  

Sea-Bird Electronics SBE 37-SMP-ODO MicroCAT  

CTD + DO Recorder

 

Quick Start Guide

 

Release Date: 09/28/2015

 

Quick Start version 

Firmware versions 

 

Software version 

#001 

SeaFET v 3.6.3 and later 

MicroCAT v 2.0 and later 

SeaFETCom V 1.2.4 and later 

Содержание SeapHOx

Страница 1: ...ed with Sea Bird Electronics SBE 37 SMP ODO MicroCAT CTD DO Recorder Quick Start Guide Release Date 09 28 2015 Quick Start version Firmware versions Software version 001 SeaFET v 3 6 3 and later MicroCAT v 2 0 and later SeaFETCom V 1 2 4 and later ...

Страница 2: ...ntents Section Page 1 Introduction 3 2 Unpacking SeapHOx 3 3 Preparing SeapHOx for Deployment 4 4 Recovering SeapHOx 5 5 Instrument Configuration reference 6 6 Cabling reference 6 7 FAQs Troubleshooting reference 7 ...

Страница 3: ...MicroCAT runs its pump before taking a measurement it takes approximately 40 seconds after a sampling event starts before the MicroCAT can output data for the SeaFET to use in the pH calculation Therefore SeaFET output frames will not appear at the telemetry port until 40 seconds have elapsed Also note that power consumption in the SeaFET is increased due to the extended time the SeaFET needs to b...

Страница 4: ...r shipment 2 If necessary configure the MicroCAT for compatibility with the SeaFET Note When delivered as part of a SeapHOx the MicroCAT is configured at the factory for integration with the SeaFET If you replace the MicroCAT in the field see Instrument Configuration at the end of this document for the required MicroCAT settings and see the MicroCAT manual for instructions on using Seaterm V2 to p...

Страница 5: ...tion in the MicroCAT manual Recovering SeapHOx 1 If the SeapHOx is not going to be used in the short term de activate the SeaFET battery pack to preserve their charge 2 Remove the SeaFET flow cell uninstall the fittings and reinstall the hex plug that is closest to the MicroCAT the one shown on the right in Figure 2 3 If needed clean the SeaFET sensors following instructions in the SeaFET manual D...

Страница 6: ...tFormat 2 Enable temperature output OutputTemp Y Set temperature units to C SetTempUnits 0 Enable salinity output OutputSal Y Enable oxygen output OutputOx Y Set oxygen units to mg l SetOxUnits 1 Enable pressure output OutputPress Y Set pressure units to dbar SetPressUnits 0 Set pumping parameters AdaptivePumpControl N OxNTau 7 Note With a typical oxygen sensor response time of OxTau20 5 5 seconds...

Страница 7: ...option allows for the pH measurement to stabilize after a new sample of water is pumped in Based on the typically observed response time the delay is factory set to 30 seconds Decreasing the delay may improve deployment duration but with a loss of accuracy Our typical recommendation is to not change the delay from the factory set value How can I verify that both instruments are working To verify t...