background image

 

Sauter GmbH

 

Ziegelei 1 
D-72336 Balingen 
E-Mail: [email protected]

 

Tel:  +49-[0]7433- 9933-199 
Fax: +49-[0]7433-9933-149 
Internet: www.kern-sohn.com 

 

 

Instruction Manual 

TC 

 

TC-BA-e-1310 

 

1

 

Digital coating thickness gauge 

 

Model: TC 1250-0.1 F 
             TC 1250-0.1 N

 

             

TC 1250-0.1 FN

 

  

Table of contents

 

 
Features 
Specifications 
Front panel description 
Measuring procedure 
Calibration 
Battery replacement 
Trouble shooting 
Restore factory settings 
Notes 
Declaration of conformity 
 

Annotation: It is strongly recommended to calibrate  
                     the new instrument before the first use, as 
                     described in paragraph 5. By doing this it  
                     will be achieved a much better measure- 
                     ment result right from the start. 

 

1. 

Features 

 

* This instrument meets the standards of both, ISO 2360  
   and DIN as well as ASTM and BS. 
   It is suitable for the laboratory and for use in `harsh field` 
   conditions. 
* The N- mode measures the thickness of nonmagnetic  
   coatings, e.g. anodizing, varnish, paint, enamel, plastic 
   coatings, powder etc. These layers are located on non- 
   magnetic metals e.g. aluminium, brass, nonmagnetic 
   stainless steel etc. 
* Automatic substrate recognition 
* Manual or automatically “auto power off” to conserve  
  batteries. 
* Two measurement modes: - single and continuous 
* Wide measuring range and high resolution 
* Date transfer to PC possible 
 

 2. Specifications 
 

Display: 4 digits 
 
Range: 0 to 1250 µm/ 0 to 50 mil 

             (another range may be specified) 
 
Resolution: 0.1 µm (0 to 100 µm) 
                   1 µm  (over 100 µm) 
Accuracy: 
- Standard:   3% of the measured value or min. ± 2.5 µm 
Is valid within a tolerance range of ± 100 

μm around the 

individually measured range, if a two-point calibration was 
performed within this tolerance range. 
 

Off-Set Accur

 Mode:   1% of the measured value 

                                      or min. ± 1.0 µm    
 
   Is valid within ± 50 

μm around the 

Off-Set Accur

 point.  

 
PC- interface: with RS-232C interface 
 
Power supply: 4x 1.5V AAA (UM-4) 
 
Operating conditions: 
                                   Temperature:         0 to 50°C 
                                   Humidity:                < 80% 
Size: 126 x 65 x 27 mm (5.0 x 2.6 x 1.1 inch) 
 
Weight: about 81g (not including batteries) 
 
Accessories: carrying case 
                     - Operation manual 
                      According to Models: 
                     1.incorporate F sensor at model TC1250-0.1F 
                      with button F/N 
                     2.  incorporate N-sensor at model  
                       TC1250-0.1N with button F/N          
                     3. incorporate FN-sensor at model  
                      TC1250-0.1 FN, Without button F/N and S/C                      
                      - Calibration foils (at each model) 
                      - Base plate (aluminium) at model N 
                      - Base plate  (iron) at model F 
                      - 2 Base plates  (iron and aluminium) for  
                        Model FN, the combination model 
 
Optional accessories: Cable & software for RS-232C 
 

3. Front panel description          

    Model TC F and TC N            Model  TC FN 

 

3-1 Sensor: F, N or FN 
3-2 Display 
3-3 Zero- key 
3-4 Plus- key 