HMO-BA-e-2020
15
5.3 Calibration
Test Calibration
The method is used to calibrate the
instrument to reduce the extent of the
error in the future.
To do this, select the "Test calibration"
function. Select the type of impact device.
Then select the scale on which the
calibration is to be carried out.
If HL is selected, the calibration mask can
be selected directly.
If a different scale is selected, select the fabric type for the test piece that is used by
default.
Perform 5 effective measurements on the test block (if the alarm function is active (ON)
and the number of strokes is less than 5, press the OK key; a query is made as to
whether operating errors should be displayed), measured values are displayed (it is
possible to delete incorrect values and return).
Touch calibration
This procedure
is used to calibrate the
correspondence between the sensor of the
touch display and the display parameters.
If there are deviations between the actual
position of press contacts of the touch display
and the actuating position of displayed values,
a touch calibration must be carried out.
To do this, simply press the centre of the
crossing line or the square frame according to
the displayed note.
Fig. 4-3
Fig. 4-1
Fig. 4-4