30 DAYS AFTER THE EXPIRY OF THE INSTRUMENT CALIBRATION, IF THE FACTORY OR SERVICE CENTRE
HAS ACTIVATED THE INSTRUMENT, A NOTIFICATION REMINDING THE USER TO SEND THE
INSTRUMENT TO THE SERVICE CENTRE WILL APPEAR ON THE DISPLAY.
Example:
15/01/14
10:00
15/01/14
10:00
15/01/14
10:00
Password
Reminder
Password
Reminder
Reminder
Calibration
Annual calibration reminder
Expiration date: 15/07/16
OK
1 1 1 1
0 0 0
F1: Info service
F2: Ignore
F3: Ignore forever
Maintain
the
pressure for
several
seconds
Enter the password
for the menu
(reminder:
1111
)
OK
OK
F1
F2
F3
7
USING THE FLUE GAS ANALYSER
7.1
Preliminary operations
Remove the analyser from the packaging used for shipping and perform an initial check of it. Check the contents match the order
specification. If you notice any signs of anomalies or damage, report the details as soon as possible to Sauermann or their representative
agent and keep the original packaging. On the label applied to the back of the device is the serial number (personnel number) of the
analyser.
This number should always be provided when there is a need for a technical intervention, spare parts, a
technical query or other clarifications.
The archive with the historical data relating to each analyser is constantly updated at Sauermann headquarters. Before using the
instrument for the first time, it is recommended that a complete battery charging cycle is carried out.
7.2
Warning
•
Use the instrument at an ambient temperature between -5 and +45°C.
•
When it has finished being used, before turning the instrument off remove the probe and let is aspirate ambient clean air for at least 30 seconds to
purge the pneumatic path from all traces of gas.
•
Do not use the instrument if the filters are clogged or damp.
•
Before putting the measuring probe back in its case after use, make sure it is has cooled down enough and there is no condensate in the tube.
•
It might be necessary to periodically disconnect the filter and the condensate separator and blow compressed air inside the tube to eliminate all
residues.
•
Remember to have the instrument checked and calibrated once a year in order to comply with the existing standards.
CONTEXT KEY
FUNCTION
F1
Displays the information about the assistance center.
F2
Ignore the notification temporarily. The next time the instrument is started, the
reminder notification will be displayed again.
F3
Permanently ignore the notification.
IN THE EVENT THAT THE INSTRUMENT HAS REMAINED AT A VERY LOW TEMPERATURE (BELOW THE
OPERATING LIMITS), IT IS SUGGESTED TO WAIT A SHORT TIME, SUCH AS 1 HOUR, BEFORE SWITCHING IT
ON. THIS WILL PROMOTE THE THERMAL BALANCE OF THE SYSTEM AND AVOID THE FORMATION OF
CONDENSATION IN THE PNEUMATIC CIRCUIT.