Measurement and accuracy ranges
MEASUREMENT
MEASURING
COMPONENT
MEASUREMENT
RANGE
RESOLUTION
DETAIL
O
2
Electrochemical sensor
0 to 25.0% vol
0.1% vol
±0.2% vol
CO
with H
2
compensation
Electrochemical sensor
0 to 8000 ppm
1 ppm
±10 ppm
0 to 200 ppm
±5% measured value
201 to 2000 ppm
±10% measured value
2001 to 8000 ppm
diluted
Electrochemical sensor
10.00% vol
0.01% vol
±20% measured value
CO low range with
H
2
compensation
Electrochemical sensor
0 to 1000 ppm
0.1 ppm
±2 ppm
0 to 40.0 ppm
±5% measured value
40.1 to 1000.0 ppm
diluted
Electrochemical sensor
6250 ppm
10 ppm
±20% measured value
CO
Electrochemical sensor
0 to 20000 ppm
1 ppm
±100 ppm
0 to 2000 ppm
±5% measured value
2001 to 4000 ppm
±10% measured value 4001 to 20000 ppm
diluted
Electrochemical sensor
25% vol
0.01% vol
±20% measured value
CO
Electrochemical sensor
0 to 10.00% vol
0.01% vol
±0.1% vol
0 to 2.00%
±5% measured value
2.01 to 10.00%
NO
Electrochemical sensor
0 to 5000 ppm
1 ppm
±5 ppm
0 to 100 ppm
±5% measured value
101 to 5000 ppm
NO low range
Electrochemical sensor
0 to 500 ppm
0.1 ppm
±2 ppm
0 to 40.0 ppm
±5% measured value
40.1 to 500.0 ppm
NOx
Calculated
-
-
-
SO
2
Electrochemical sensor
0 to 5000 ppm
1 ppm
±5 ppm
0 to 100 ppm
±5% measured value
101 to 5000 ppm
SO
2
low range
Electrochemical sensor
0 to 500 ppm
0.1 ppm
±2 ppm
0 to 40.0 ppm
±5% measured value
40.1 to 500.0 ppm
NO
2
Electrochemical sensor
0 to 1000 ppm
1 ppm
±5 ppm
0 to 100 ppm
±5% measured value
101 to 1000 ppm
NO
2
low range
Electrochemical sensor
0 to 500 ppm
0.1 ppm
±2 ppm
0 to 40.0 ppm
±5% measured value
40.1 to 500.0 ppm
CxHy
Pellistor sensor
0 to 5.00% vol
0.01% vol
±0.25% vol
CO
2
Calculated
0 to 99.9% vol
0.1% vol
-
CO
2
NDIR sensor
0 to 20.0% vol
0.1% vol
±0.3% vol
0.00 to 6.00%
±5% measured value
6.01 to 20.0%
H
2
S
Electrochemical sensor
0 to 500 ppm
0.1 ppm
±5 ppm <125.0 ppm
±4 % of the reading <500.0 ppm
PI*
(CO/CO
2
report)
Calculated
-
0.01%
-
Air temperature
TcK sensor
-20.0 to 1250.0 °C
0.1 °C
±0.5 °C
0 to 100 °C
±0.5% measured value
101 to 1250 °C
Flue gas temperature
TcK sensor
-20.0 to 1250.0 °C
0.1 °C
±0.5 °C
0 to 100 °C
±0.5% measured value
101 to 1250 °C
Pressure
(draught & differential)
Piezoelectric sensor
-10.00 to 200.00 hPa
0.01hPa
±1% measured value 2.01 to 10.00 hPa
±0.02 hPa
-2.00 to +2.00 hPa
±1% measured value+2.01 to +200.00 hPa
Differential temperature
Calculated
0 to 1250.0 °C
0.1 °C
-
Air index
Calculated
0.00 to 9.50
0.01
-
Excess air (“e”)
Calculated
0 to 850%
1%
-
Stack loss
Calculated
0.0 to 100.0%
0.1%
-
Efficiency
Calculated
0.0 to 100.0%
0.1%
-
Efficiency
(condensing)
Calculated
0.0 to 120.0%
0.1%
-
Smoke index
External instrument:
0 to 9
-
-
* The Poison Index (P.I.) is a reliable indicator of the proper operation of the burner or boiler. In this way, through a simple flue analysis, it can be determined
whether or not maintenance is necessary.