Satelec X-Mind DC Скачать руководство пользователя страница 1

X-Mind

®

DC

Installation manual / Manuel d’installation

Installationshandbuch / Installatiehandleiing

Содержание X-Mind DC

Страница 1: ...X Mind DC Installation manual Manuel d installation Installationshandbuch Installatiehandleiing ...

Страница 2: ...X Mind DC Intraoral X ray system at constant potential INSTALLATION MAINTENANCE MANUAL This manual should always be kept in proximity to the device ...

Страница 3: ...E FEEDING TERMINAL BOARD page 21 5 6 ASSEMBLING THE TUBEHEAD page 24 5 7 BALANCING THE PANTOGRAPH TYPE ARM page 27 5 8 ASSEMBLING THE X Mind DC TIMER page 30 5 9 TIMER ELECTRIC CONNECTION page 32 6 CONTROL PANEL page 34 7 SYSTEM CONFIGURATION page 36 8 CHANGING THE CONFIGURATION page 38 9 START UP page 40 10 CHECKING THE INSTALLATION page 41 11 CHECKING THE EXPOSURE FACTORS page 43 12 DIAGNOSTICS ...

Страница 4: ...NING and PLEASE NOTE messages when operating the system LEGEND CAUTION The word CAUTION refers to events which might compromise the operator s personal safety or cause injuries to people WARNING The word WARNING refers to events which might compromise the radiographic system s performance PLEASE NOTE The words PLEASE NOTE refer to special indications to facilitate maintenance or make important inf...

Страница 5: ...age mentioned in the rating plate matches the line voltage Install the X Mind DC X ray system according to the procedures described in this manual Provide the user with any information regarding the use of the X Mind DC X ray system pursuant to the manual Certify the work done by a Statement of Compliance Return the duly filled in warranty certificate to Satelec S A S if this is not done the warra...

Страница 6: ...GRAPH TYPE ARM TIMER WALL PLATE BRACKET OPTIONAL long 12 31 cm cone short 8 20 cm cone long cone 12 31 cm with a rectangular section sized 44x35mm second CONTROL BUTTON with extension cable RX signaling lamp for external use X Mind LIGHT remote control button X Mind ECB Fig 1 1 X Mind DC X RAY SYSTEM ...

Страница 7: ...D TAG OF THE LONG 12 CONE ID TAG OF THE SHORT 8 CONE ID TAG OF THE RECTANGULAR CONE GRADUATED SCALE TAG ON TUBEHEAD Pictograms used This symbol guarantees that the X ray system complies with the regulations contained in the European Directive EEC 93 42 regarding Medical Devices The degree of protection against direct and indirect electric contacts is B type Refer to Manual s instructions N NEUTRAL...

Страница 8: ...sions of the possible supply conditions SHORT BRACKET optional length 41 cm 16 2 STANDARD BRACKET length 82 5 cm 32 5 LONG BRACKET optional length 110 cm 43 5 SHORT BRACKET 41 cm Fig 1A STANDARD BRACKET 82 5 cm Fig 1B LONG BRACKET 110 cm Fig 1C 3 OVERALL DIMENSIONS ...

Страница 9: ...8 8 X X M Mi in nd d D DC C Fig 2 3A 3B 3C e 4 show the typical dimensions of the X ray system STANDARD BRACKET 82 5 cm Fig 2 ...

Страница 10: ...X X M Mi in nd d D DC C 9 9 STANDARD BRACKET 82 5 cm Fig 3A Fig 3B Fig 3C SHORT BRACKET 41 cm LONG BRACKET 110 cm ...

Страница 11: ...1 10 0 X X M Mi in nd d D DC C Fig 4 ...

Страница 12: ... 50 C e The relative humidity must range within 25 and 75 2 SUPPORTING WALL REQUIREMENTS a The X ray system supporting wall must be able to stand 200 kg tear at every point PLEASE NOTE The nature and consistency of the wall must be checked and if required seek the support of a brickwork expert Walls of uncertain consistency must be reinforced with a buried counter plate or with a sandwich type sys...

Страница 13: ...ONDUCTOR SECTION MAXIMUM LINE LENGTH 1 5 mm2 10 meters 196V to 264V 2 5 mm2 20 meters 1 5 mm2 10 meters 98V to 132V 2 5 mm2 20 meters PLEASE NOTE For longer lines the cord section must be increased in proportion d The communication cables C11 C12 C21 C22 between the timer and X Ray must be two pole twisted and shielded with 0 25 mm2 section example Belden 9501 type e The cables S11 S12 S21 S22 con...

Страница 14: ... be provided pursuant to the electrical installation diagram refer to 20 ATTACHMENTS a Timer electrical cables refer to 5 9 TIMER ELECTRIC CONNECTIONS b Cables connecting the timer and the tubehead refer to 5 9 TIMER ELECTRIC CONNECTIONS c Cables connecting the timer and the RX signal lamp s for external use X Mind LIGHT OPTIONAL d Cables connecting the timer and the remote control button X Mind E...

Страница 15: ... the following sketch Fig 5 D DOCUMENTS INSTRUCTION MANUAL INSTALLATION MAINTENANCE MANUAL AND WARRANTY CARD X Mind DC TUBEHEAD PACKAGING X Mind DC TIMER RX SIGNALING LAMP FOR EXTERNAL USE X Mind LIGHT OPTIONAL REMOTE CONTROL BUTTON X Mind ECB OPTIONAL PACKAGING PANTOGRAPH TYPE ARM PACKAGING WALL PLATE BRACKET PACKAGING PLEASE NOTE Prior to installation duly check all components PLEASE NOTE The ca...

Страница 16: ...e tubehead PLEASE NOTE To prevent any flaking in the white coat and to control the center distances between the holes it is advisable to start drilling with a Ø7 bit increasing this size gradually 7 Drill the wall plate anchor holes 8 If required drill the holes for the electric cables connecting the timer to the tubehead 9 Remove the template and insert the suitable anchor screws according to the...

Страница 17: ...1 16 6 X X M Mi in nd d D DC C Fig 6 ...

Страница 18: ...stop key Fig 7A and 7B to prevent the electrical cable from twisting PLEASE NOTE Generally the stop key is installed so that the equipment position at rest is on the right side viewed from the front of the wall plate Fig 7A Should the position at rest be on the left side the stop key must be rotated by 180 Fig 7B 5 3 ASSEMBLING THE BRACKET Fig 7A Fig 7B WALL ...

Страница 19: ...5 2 Insert the bracket pin into the wall plate upwards 3 Insert the supporting rest Fig 8 PLEASE NOTE Prevent all foreign matter ground dust cement etc from settling on the pin seat The pin must slide freely in its seat If required thoroughly clean and lubricate with molikote D lubricant PLEASE NOTE Carefully check that the bracket is parallel to the floor using a water level ...

Страница 20: ...rd wafer 4 Insert the pantograph group cable into the washer and then the pantograph pin Fig 9 5 If required clean the pin and the washer and lubricate with molikote D lubricant 6 Insert the electric cable into the bracket housing 7 Assemble the guard wafer 8 Insert the cable into the bracket and push it until it reaches the pin outlet near the supply terminal board 5 4 ASSEMBLING THE PANTOGRAPH T...

Страница 21: ...le and of the communication cable is suitable for 82 5 cm brackets DOT NOT tamper with the terminals arrangement when 41 cm long brackets are used The excess cable must be stored in the housing provided WARNING In case of electric power supply type PHASE PHASE the higher value must be taken using a tester and entered in L 5 5 CONNECTION TO THE FEEDING TERMINAL BOARD ...

Страница 22: ...X X M Mi in nd d D DC C 2 21 1 OPERATING INSTRUCTIONS Fig 10 1 Remove the terminal board cover by unscrewing the set screw Fig 10 ...

Страница 23: ...UE N neutral BLACK COMMUNICATIONS RED COMMUNICATIONS YELLOW GREEN T ground YELLOW GREEN T ground THE COMMUNICATIONS CABLES ARE NOT POLARIZED 3 Connect the pantograph cable shield to the grounding potential 4 Connect the wall plate to the grounding potential 5 Clamp the cables with the cable clamps provided 6 Reassemble the terminal board cover refer to Fig 10 IMPORTANT ...

Страница 24: ... to Fig 5 2 Check that all the rating data match the power supply voltage 3 Remove both guards from the pantograph type arm by loosening the appropriate screws Fig 12 Fig 12 4 Using a bit detach the front coupling device Fig 13A 5 Remove both guards Fig 13B Fig 13A Fig 13B 5 6 ASSEMBLING THE TUBEHEAD ...

Страница 25: ...tubehead pin into the pantograph head Fig 14A and insert the support Fig 14B Fig 14A Fig 14B 7 Check that during insertion the pin of the rotation prevention device is correctly inserted into the seat located on the pantograph head Fig 15 Fig 15 ...

Страница 26: ...X X M Mi in nd d D DC C 2 25 5 8 Couple the pantograph and tubehead connectors and insert them into their seats Fig 16 Fig 16 ...

Страница 27: ... balancing tests the adjustment key must not be left in place WARNING The key provided must be carefully stored PLEASE NOTE To reach the X adjustment screw ARM A must be put in vertical position To reach the Y adjustment screw ARM B must be put in horizontal position The adjustment key provided can be inserted only under the above conditions Fig 17 5 7 BALANCING THE PANTOGRAPH TYPE ARM ARM B Y ARM...

Страница 28: ...he correct position 1 BRING ARM B BACK TO THE HORIZONTAL POSITION 2 INSERT THE ADJUSTMENT KEY IN Y 3 ROTATE THE ADJUSTMENT KEY BY HALF TURN CLOCKWISE IF IT TENDS TO COME DOWN COUNTERCLOCKWISE IF IT TENDS TO GO UP 4 WITHDRAW THE KEY PLEASE NOTE Repeat the test and adjustment until ARM B is steady and stable in all positions even when ARM A is completely extended 4 READJUSTMENT OF ARM A If it does n...

Страница 29: ...lancing 1 Insert the movable finish A between the pantograph type arm guard and the metal frame Fig 18 Fig 18 2 Insert the pins of the guards into their slots then position them and ensure that the movable finish is coupled to the guards Fig 19 Fig 19 A ...

Страница 30: ...data match the power supply voltage ASSEMBLY INSTRUCTIONS Fig 20 1 Take the timer out of the packaging refer to Fig 5 2 Take out the drilling template 3 Mark the anchor holes on the wall using the drilling template 4 Drill using a Ø3mm bit then drill again with a Ø6 mm bit to prevent any flaking of the white coat 5 Insert the three screw anchors provided into the holes 6 Open the timer by unscrewi...

Страница 31: ...t the cables of the remote control button IF INSTALLED Fig 22 12 Apply the timer base to the wall matching the three anchor screws with the holes 13 Insert and tighten the screws refer to Fig 20 with the appropriate washers refer to Fig 20 into the screw anchors PLEASE NOTE If the wall is uneven insert some shims to prevent the timer from buckling PLEASE NOTE Do not contaminate the timer with dust...

Страница 32: ...k 3 Fix them with the cable clamp 4 Connect the cables coming from tubehead n 1 to the terminals XRAY1 5 Connect the YELLOW GREEN grounding cable to the equipotential metal plate 6 Connect the cables coming from tubehead n 2 to terminals XRAY2 7 Connect the YELLOW GREEN grounding cable to the equipotential metal plate 8 Clamp the cables in the cable clamp 9 Connect the RX signaling lamps for exter...

Страница 33: ...TTON 3A 3A TUBEHEAD N 1 RX SIGNALING LAMP FOR EXTERNAL USE 3B 3B TUBEHEAD N 2 RX SIGNALING LAMP FOR EXTERNAL USE The electronic card terminal boards control the following functions 4A 4A RS232 COMMUNICATION FOR XRAY1 4B 4B RS232 COMMUNICATION FOR XRAY2 5 5 TUBEHEAD N 1 POWER SUPPLY 6 6 TUBEHEAD N 2 POWER SUPPLY 7 7 TIMER POWER SUPPLY 1 2 3A 3B 4A 4B 5 6 7 ...

Страница 34: ... s interface from which the operator can use the keys to select the operating parameters to change the exposure times displayed to store the new customized exposure times use the green LED light indicators to check the timer operating conditions 6 CONTROL PANEL ...

Страница 35: ...D TYPE INDICATOR X RAY DISTANCE INDICATOR TUBEHEAD SELECTION X RAY VOLTAGE INDICATOR MEMORY SELECTION OF PATIENT S PHYSIQUE TYPE UPPER MAXILLARY TEETH LOWER MANDIBLE TEETH X RAY CURRENT INDICATOR X RAY OUTPUT SIGNAL OCCLUSAL EXAM BITEWING EXAM PAUSE INDICATOR MALFUNCTION INDICATOR DIGITAL X RAY TECHNIQUE CONVENTIONAL X RAY TECHNIQUE KEY SWITCH ...

Страница 36: ...led tube head has the following features work with direct current at constant potential the DC LED on the control panel must be lit work with an X ray voltage equal to 70kVp the 70 kVp LED on the control panel must be lit work with an X ray current equal to 8mA the 8 mA LED on the control panel must be lit The above configuration depends on the position of 8 mini switches dip switch on the timer e...

Страница 37: ...button 6 8 cone 20 cm 7 not available 8 not available The configuration may be changed if POSSIBLE MODIFICATION HOW TO CARRY OUT THE MODIFICATIONS type E and F films are used a digital system is used use of a voltage equal to 60 kV use of a current equal to 4 mA refer to USER S MANUAL 6 OPERATING INSTRUCTIONS the short 8 20 cm cone is used the long 12 31 cm cone is used a single tubehead is used u...

Страница 38: ...METER ON OFF 1 TUBEHEAD No 1 INSTALLED NOT INSTALLED 2 TYPE OF TUBEHEAD No 1 dc DIRECT CURRENT AT CONSTANT POTENTIAL ac ALTERNATING CURRENT SINGLE PHASE 3 TUBEHEAD No 2 INSTALLED NOT INSTALLED 4 TYPE OF TUBEHEAD No 2 dc DIRECT CURRENT AT CONSTANT POTENTIAL ac ALTERNATING CURRENT SINGLE PHASE 5 No 2 CONTROL BUTTON INSTALLED NOT INSTALLED 6 CONE LONG 12 SHORT 8 7 NOT AVAILABLE NOT AVAILABLE 8 NOT AV...

Страница 39: ...nge the type of tubehead move dip switch no 2 or no 4 a If tubehead RX1 is ac put dip switch no 2 in the OFF position otherwise put it in the ON position b If tub head RX2 is ac put dip switch no 4 in the OFF position otherwise put it in the ON position 3 CHANGING THE CONE To replace the long 12 cone with the short 8 cone move dip switch no 6 to the OFF position Check that the 8 LED lights up on t...

Страница 40: ...f an error is detected when the system is turned on the anomaly is indicated as follows an intermittent beep sounds the MALFUNCTION INDICATOR LED blinks intermittently The error code E appears on the display refer to 13 ERROR MESSAGES All control panel functions are disabled In this event turn off the timer and then turn it back on If the error should recur call the Support Department PLEASE NOTE ...

Страница 41: ...ime of 1 sec b Take the timer s CONTROL BUTTON c Using the extension cable of the CONTROL BUTTON maintain a safe distance of at least 2 meters from the tube head so as to be able to constantly check the X ray exposure d Press the X RAY key and keep it depressed until the acoustic signal beep stops and the yellow X RAY OUTPUT SIGNAL LED turns off e At the end of the exposure the green PAUSE INDICAT...

Страница 42: ...ne voltage 15 Otherwise check the electrical system or call the Support Department 6 CHECKING THE ELECTRIC SYSTEM To check the electric system a tester must be used in the mode voltmeter in AC a Connect the instrument to the L and N terminals on the timer b Measure the line voltage V0 c Connect the instrument to the L and N terminals of the wall plate terminal board refer to 5 5 CONNECTION TO THE ...

Страница 43: ... 8mA 10 3 DOSE mGy The dose in the air is measured with a non invasive instrument by positioning the detector at an SSD source skin distance of 31 cm 12 or 20 cm 8 Line Voltage Nominal V 15 Max Voltage Drop 3 kVp 60 70 mA 4 8 SET TECHNICAL VALUES Set Exposure Time 1 sec DOSE in the air SSD SOURCE SKIN DISTANCE 31 CM 12 SSD SOURCE SKIN DISTANCE 20 CM 8 70 kVp 8 mA 5 5 mGy s 30 70 kVp 8 mA 12 mGy s ...

Страница 44: ...4 the message SEC is displayed for approx 1 sec 5 the current value of the lower set limit is displayed 6 to change the value press the keys and 7 to confirm press the key e To exit this mode turn the timer off and then on again 2 To display the parameters the installer must a Simultaneously press and keep depressed the 17 UPPER MAXILLARY MOLAR 47 LOWER MAXILLARY MOLAR keys b Press the key associa...

Страница 45: ...e timer b Switch on the timer and keep the mA button depressed c The display shows the tube message d Select X RAY with the button to calibrate e Take the exposure Press and keep the RX button depressed until the beeping stops stops the yellow X RAY OUTPUT SIGNAL LED goes off f Once the exposure has been taken if the display does not show ERRORS the calibration has been successfully completed othe...

Страница 46: ... NOT INCLUDED WITHIN THE 15 NOMINAL VALUE CALL THE SUPPORT DEPARTMENT THE X RAY KEY ALWAYS SEEMS TO BE DEPRESSED ENSURE THAT IT IS NOT JAMMED ANOMALY ON THE CONTROL PANEL CALL THE SUPPORT DEPARTMENT THE EXPOSURE HAS BEEN PREMATURELY INTERRUPTED KEEP THE X RAY KEY DEPRESSED TILL THE END OF THE EXPOSURE ANOMALY IN THE TRIAC RELAY CALL THE SUPPORT DEPARTMENT ANOMALY IN THE ELECTRONIC CIRCUIT CALL THE...

Страница 47: ...ENT EEPROM DATA NOT PROPERLY SAVED CALL THE SUPPORT DEPARTMENT CORRUPTED EEPROM DATA CALL THE SUPPORT DEPARTMENT OVERVOLTAGE ERROR CALL THE SUPPORT DEPARTMENT ANODE VOLTAGE OUT OF TOLERANCE CALL THE SUPPORT DEPARTMENT ANODE CURRENT OUT OF TOLERANCE CALL THE SUPPORT DEPARTMENT CONTROL CONNECTOR CALL THE SUPPORT DEPARTMENT REFERENCE VOLTAGE PROBLEM CALL THE SUPPORT DEPARTMENT MAJOR ERROR ALL X RAY S...

Страница 48: ... remove the timer s guard by removing the adjusting 3 Determine the fuse to be replaced Fig 26 4 Remove the plastic protector Fig 27 5 Remove the fuse 6 Replace it with a fuse of same type POWER VOLTAGE OF THE X RAY SYSTEM 230 V 15 115 V 15 Internal protection fuses 5x20 quick 8AF 250V 12 5AF 250V 7 Insert the protection Fig 26 Fig 27 8 Close the timer s guard 9 Restore the power 15 FUSE REPLACEME...

Страница 49: ...e bracket shaft use molikote D lubricant 16 Lubricate the bracket bush with lubrication grease use molikote D lubricant 17 Install the bracket in the wall plate refer to 5 3 ASSEMBLING THE BRACKET 18 Check the pantograph type arm cable If it is be damaged send the pantograph type arm to the manufacturer for repairs 19 Check the pantograph type arm guards 20 Replace the damaged guards 21 Remove all...

Страница 50: ... S Z I du Phare B P 216 33708 MERIGNAC cedex France Tel 33 0 556 34 06 07 Fax 33 0 556 34 92 92 E mail info satelec com In the event of disposal the components and packaging must be disposed of in an environmentally correct manner In particular the dielectric oil as well as the shielding lead must be disposed by means of authorized companies specialized in the disposal of waste material 17 CLEANIN...

Страница 51: ...kes to supply upon request drawings circuit diagrams component parts lists instructions or other information needed by qualified technical personnel to perform repairs on those parts of the X Mind DC radiographic system which may be repaired 20 ACCESSORIES ...

Страница 52: ...X X M Mi in nd d D DC C 5 51 1 X Mind DC Installation electrical scheme All rights reserved ...

Страница 53: ...FORCELLA 29700980 5 3 AC DC XG 3 FORCELLA 29700777 3 4 AC DC XG 4 LIMITATORE DI ROTAZIONE 29700750 1 5 DC XG 5 GUARNIZIONE 27500959 1 6 DC XG 6 SCHEDE ARIA 29700981 9 7 DC XG 7 COPERTURA SCHEDA ARIA 29700982 11 8 DC XG 8 CAVO TESTA 39700385 1 A CONO LUNGO 31cm 12 Ø60 29700397 1 B CONO CORTO 20cm 8 Ø60 29700396 1 C CONO QUADRO 31cm 12 44x35 29700615 1 ...

Страница 54: ... LIGHT CODE Q TY PCS 1 XG LIGHT 1 GUSCIO INF SUP 29700977 4 2 XG LIGHT 2 PORTALAMPADA 33400372 1 3 XG LIGHT 3 LAMPADINA 230V 15W 33300352 1 3 XG LIGHT 3 LAMPADINA 115V 15W 33300390 1 4 XG LIGHT 4 SCHEDA 230V 39200354 3 4 XG LIGHT 4 SCHEDA 115V 39200353 3 ...

Страница 55: ...5 54 4 X X M Mi in nd d D DC C REF SPARE PARTS X MIND ECB CODE Q TY PCS 1 XG ECB 1 GUSCIO INF SUP 29700978 4 2 SCHEDA COMANDO CON PULSANTE E LED 39700303 1 3 XG ECB 3 INTERRUTTORE A CHIAVE 35100298 1 ...

Страница 56: ... 32000359 1 5 AC DC XG TIMER 5 CASSA PULSANTE 29700974 4 6 SCHEDA COMANDO CON PULSANTE E LED 39700303 1 7 AC DC XG TIMER 7 SCHEDA LED 39200313 7 8 AC DC XG TIMER 8 TASTIERA DEG 39200308 1 8 AC DC XG TIMER 8 TASTIERA SAT 39200309 1 9 AC DC XG TIMER 9 CASSA 29700975 1 10 AC DC XG TIMER 10 COPERCHIO 29700976 4 11 AC DC XG TIMER 11 BLOCK CHIAVE 39700305 2 12 DC XG TIMER 12 FILTRO 36200382 3 13 INTERRU...

Страница 57: ...0460 4 5 BRACCIO PANTOGRAFO PORTATIRANTE 29700458 6 6 AC DC XG B 6 TAPPI set 29700991 7 7 AC DC XG B 7 COPERTURA EST 27500757 1 8 AC DC XG B 8 COPERTURA INT 27500758 1 9 AC DC XG B 9 GUSCI CENTRALI 29700993 8 10 BRACCIO PANTOGRAFO CHIAVE CARICAMENTO 22000543 1 11 AC DC XG B 11 COPERTURA INT 27500758 1 12 AC DC B 12 COPERTURA EST 27500757 1 13 AC DC XG B 13 TIRANTE POST 29700992 7 14 BRACCIO PANTOG...

Страница 58: ...ALL PLATE 350 CODE Q TY PCS 1 PIASTRA 1 COPERTURA 29700985 5 1 AC DC XG PIASTRA 1 COPERTURA SAT 29700986 5 2 PIASTRA LUNETTA 22000176 1 3 PIASTRA 3 DIMA 24200984 1 PIASTRA MURO 350 DEG 29700383 COMPLETO AC DC XG PIASTRA MURO 350 SAT 29700987 COMPLETO ...

Страница 59: ...E 350 TOP CODE Q TY PCS 1 PIASTRA 1 COPERTURA 29700985 5 1 AC DC XG PIASTRA 1 COPERTURA SAT 29700986 5 2 PIASTRA LUNETTA 22000176 1 3 PIASTRA 3 DIMA TOP 24201002 1 PIASTRA MURO 350 TOP DEG 29700988 COMPLETO AC DC XG PIASTRA MURO 350 TOP SAT 29700989 COMPLETO ...

Страница 60: ...29700379 COMPLETO REF SPARE PARTS CEILING BRACKET CODE Q TY PCS MENSOLA PLAFONE 29700382 COMPLETO REF SPARE PARTS UNIT BRACKET CODE Q TY PCS MENSOLA RIUNITO 29700381 COMPLETO REF SPARE PARTS TOP MOUNT BRACKET CODE Q TY PCS MENSOLA 400 TOP MOUNT 29700620 COMPLETO MENSOLA 800 TOP MOUNT 29700621 COMPLETO MENSOLA 1100 TOP MOUNT 29700622 COMPLETO ...

Страница 61: ...4 5 BRACCIO PANTOGRAFO PORTATIRANTE 29700458 6 6 AC DC XG B 6 TAPPI set 29700991 7 7 AC DC XG B 7 COPERTURA EST 27500757 1 8 AC DC XG B 8 COPERTURA INT 27500758 1 9 AC DC XG B 9 GUSCI CENTRALI 29700993 8 10 BRACCIO PANTOGRAFO CHIAVE CARICAMENTO 22000543 1 11 AC DC XG B 11 COPERTURA INT 27500758 1 12 AC DC B 12 COPERTURA EST 27500757 1 13 AC DC XG B 13 TIRANTE POST 29700992 7 14 BRACCIO PANTOGRAFO ...

Страница 62: ... MUURPLAAT 350 CODE AANTAL 1 PIASTRA 1 COPERTURA 29700985 5 1 AC DC XG PIASTRA 1 COPERTURA SAT 29700986 5 2 PIASTRA LUNETTA 22000176 1 3 PIASTRA 3 DIMA 24200984 1 PIASTRA MURO 350 DEG 29700383 COMPLETO AC DC XG PIASTRA MURO 350 SAT 29700987 COMPLETO ...

Страница 63: ...AT 350 TOP CODE AANTAL 1 PIASTRA 1 COPERTURA 29700985 5 1 AC DC XG PIASTRA 1 COPERTURA SAT 29700986 5 2 PIASTRA LUNETTA 22000176 1 3 PIASTRA 3 DIMA TOP 24201002 1 PIASTRA MURO 350 TOP DEG 29700988 COMPLETO AC DC XG PIASTRA MURO 350 TOP SAT 29700989 COMPLETO ...

Страница 64: ...00 29700379 COMPLETO REF WISSELSTUKKEN PLAFONDBEUGEL CODE AANTAL MENSOLA PLAFONE 29700382 COMPLETO REF WISSELSTUKKEN EENHEIDBEUGEL CODE AANTAL MENSOLA RIUNITO 29700381 COMPLETO REF WISSELSTUKKEN BOVENBEUGEL CODE AANTAL MENSOLA 400 TOP MOUNT 29700620 COMPLETO MENSOLA 800 TOP MOUNT 29700621 COMPLETO MENSOLA 1100 TOP MOUNT 29700622 COMPLETO ...

Страница 65: ... 44 1480 477 381 e mail info uk acteongroup com MIDDLE EAST ACTEON MIDDLE EAST Numan Center 1st Floor N 111 Gardens Street PO Box 468 AMMAN 11953 JORDAN Tel 962 6 553 4401 Fax 962 6 553 7833 e mail info me acteongroup com CHINA ACTEON CHINA Office 401 12 Xinyuanxili Zhong Street Chaoyang District BEIJING 100027 CHINA Tel 86 10 646 570 11 2 3 Fax 86 10 646 580 15 e mail info cn acteongroup com PHIL...

Отзывы: