The YDI01C-WP digital interface built into
the Combics indicator enables connection
of any of the following:

– Digital weighing platform 

(e.g. IS weighing platform) using the
YCC02-R12F6 connecting cable (Option
M41: 12-contact round female connector;
length: 6 m; open cable ends), or

– Digital weighing instrument (e.g., LA, LP,

ME, EA, EB, CP balance/scale) using the
YCC02-D25F6 connecting cable (Option
M42: 25-pin D-Sub male connector;
length: 6 m; open cable ends), or

– Digital weighing instrument (e.g. QA, 

QC, FB, FC scale) using the YCC02-R12M6
connecting cable (Option M43: 12-pin
round male connector; length: 6 m; 
open cable ends).

Note: 

To have devices communicate with the Combics indicator as plat-

forms and scales through the SBI protocol, configure the operating menu
as follows:
Baud rate: 9600

Handshake mode: Hardware, 1 character after CTS

Parity:

Odd

Number of data bits: 7

Stop bits: 1
Printing: 

Automatic without stability

Identification:

With ID code (22 characters)

To have devices communicate with the Combics indicator as platforms and
scales over the xBPI protocol, configure the operating menu as follows:
Setting:

xBPI protocol

Interface:

RS-232 or RS-485

!

Refer to the relevant operating instructions for descriptions of the 

operating menus. 
The menu setting in the Combics indicator depends on the particular 
platform to be connected:
Setting:

SBI or xBPI protocol

Interface:

RS-232 or RS-485

Example: For Combics 1 and 2, this setting is configured in the Setup
menu; for Combics 3, in the menu under Device parameters
WP1

RS-232

or

RS-485

SBI: Standard

o IS-485

SBI: Trade

ADC-485

o IS-232

ADC-232

!

Always make sure the equip-
ment is disconnected from
power before performing any
installation, maintenance or
repair work.

!

Work that affects IP protec-
tion must be carried out with
extreme care.

!

Any installation work that
does not conform to the
instructions in this manual
will void all claims under the
manufacturer’s warranty.

!

Do not operate the YDI01C-
WP in hazardous areas. 

§

Place the Combics indicator
on a soft surface and remove
the four cap nuts from the
front panel. 

§

Remove the front panel

Installing the YDI01C-WP PCB 

§

Remove the A/D converter
PCB

– Unplug the male connector
– Remove the 3 screws 
– Remove the A/D converter

PCB 

§

Plug in the YDI01C-WP PCB 

– Use the 2 screws to fasten

the PCB

– Plug in the male connector