Installation Instructions
Sartorius
YDI01C-WP (Option
A6|A7
)
Installing a Digital Interface in Place of an A/D Converter in
Combics Indicators, Models CIS1..|CIS2..|CIS3.. | CISL1..|CISL2..|CISL3..
98647-003-78
The YDI01C-WP digital interface built into
the Combics indicator enables connection
of any of the following:
– Digital weighing platform
(e.g. IS weighing platform) using the
YCC02-R12F6 connecting cable (Option
M41: 12-contact round female connector;
length: 6 m; open cable ends), or
– Digital weighing instrument (e.g., LA, LP,
ME, EA, EB, CP balance/scale) using the
YCC02-D25F6 connecting cable (Option
M42: 25-pin D-Sub male connector;
length: 6 m; open cable ends), or
– Digital weighing instrument (e.g. QA,
QC, FB, FC scale) using the YCC02-R12M6
connecting cable (Option M43: 12-pin
round male connector; length: 6 m;
open cable ends).
Note:
To have devices communicate with the Combics indicator as plat-
forms and scales through the SBI protocol, configure the operating menu
as follows:
Baud rate: 9600
Handshake mode: Hardware, 1 character after CTS
Parity:
Odd
Number of data bits: 7
Stop bits: 1
Printing:
Automatic without stability
Identification:
With ID code (22 characters)
To have devices communicate with the Combics indicator as platforms and
scales over the xBPI protocol, configure the operating menu as follows:
Setting:
xBPI protocol
Interface:
RS-232 or RS-485
!
Refer to the relevant operating instructions for descriptions of the
operating menus.
The menu setting in the Combics indicator depends on the particular
platform to be connected:
Setting:
SBI or xBPI protocol
Interface:
RS-232 or RS-485
Example: For Combics 1 and 2, this setting is configured in the Setup
menu; for Combics 3, in the menu under Device parameters
WP1
RS-232
or
RS-485
SBI: Standard
o IS-485
SBI: Trade
ADC-485
o IS-232
ADC-232
!
Always make sure the equip-
ment is disconnected from
power before performing any
installation, maintenance or
repair work.
!
Work that affects IP protec-
tion must be carried out with
extreme care.
!
Any installation work that
does not conform to the
instructions in this manual
will void all claims under the
manufacturer’s warranty.
!
Do not operate the YDI01C-
WP in hazardous areas.
§
Place the Combics indicator
on a soft surface and remove
the four cap nuts from the
front panel.
§
Remove the front panel
Installing the YDI01C-WP PCB
§
Remove the A/D converter
PCB
– Unplug the male connector
– Remove the 3 screws
– Remove the A/D converter
PCB
§
Plug in the YDI01C-WP PCB
– Use the 2 screws to fasten
the PCB
– Plug in the male connector