Sartorius
XQ 6100, XQ 6100 B.
Electronic Paint Mixing Scales
Installation and Operating Instructions
Страница 1: ...Sartorius XQ 6100 XQ 6100 B Electronic Paint Mixing Scales Installation and Operating Instructions...
Страница 2: ...the Implemen tation of Regulations for Prevention of Accidents Elektrische Anlagen und Betriebsmittel VBG 4 Electrical Installations and Equipment of April 1986 it is hereby certified that the equipm...
Страница 3: ...3 Contents Page Equipment Supplied 5 Installation Instructions 6 Startup 7 Operation 8 Examples for Paint Mixing 9 Calibration 15 Accessories Options 17 Spezifications 17...
Страница 4: ...lay 4 TINTER ADDITIVE key 5 Set 100 key 6 Access switch for calibration 7 Tare control 8 PRINT key functions only if the scale has a built in interface 9 RECALL WEIGHT Taste 10 ON OFF key 11 g key 12...
Страница 5: ...use The equipment supplied comprises the components shown on the left Scale Dust cover AG adapter Weighing pan lf you wish to send in your scale for repair work please pro ceed as follows remove all...
Страница 6: ...connect or disconnect additional devices printer or PC to or from the interface connector Option Startup Remove the white backing from the adhesive disks on the dust cover Attach the dust cover to th...
Страница 7: ...iring warmup the next time you switch it on Important Note To use the STANDBY feature the scale must have been plugged into the AC adapter for at least 30 minutes initial warmup After the scale is tur...
Страница 8: ...small circle in the lower left hand corner shows that the scale has been exactly tared so the display reads 0 Do not wait to replace worn dust covers Make sure that paint spills over the sides of the...
Страница 9: ...al weight of the paint components and hardener Place an empty container on the scale and tare Weigh in the 1 st paint component according to your formula and then store it press the key 4 weigh in the...
Страница 10: ...isappear once the total weight is stored as 100 or E will be displayed if the paint components are too light for the particular shade of paint If E is displayed press the RECALL WEIGHT key and in crea...
Страница 11: ...ortant Note Make sure that the 20 thinner is in proportion to the total weight of the paint components and to the hardener to display the update total weight paint components hardener thinner press th...
Страница 12: ...on to the total weight of the paint quantity 1 st paint component 2nd component 100 Place empty container on the scale and tare To weigh in the paint components according to your for mula and store th...
Страница 13: ...e the total weight is stored as 100 or E will be displayed if the paint components are too light for the particular shade of paint lf E is displayed press the RECALL WEIGHT key and in crease the minim...
Страница 14: ...in percent by pressing the key 9 To display the total net weight in the particular weight unit se lected in this case g press the key 11 The key 9 deletes the stored factor for calculating the percen...
Страница 15: ...ove the protective cap located on the front right of your scale and slide the access switch 6 in the direction of the ar row Turn off the scale hold down the TINTER ADDITIVE key 4 and briefly press th...
Страница 16: ...itch back to its original setting and replace the protective cap After calibration use the ON OFF key 10 to turn the scale off and then on again Even if you do not calibrate your scale this will not a...
Страница 17: ...lity depends on the filter level selected when load is changed s s 0 1 0 2 0 4 0 1 Adaptation to operating requirement and ambient conditions by selecting one of 3 optimised filter level Stability ran...
Страница 18: ...Copyright by Sartorius AG G ttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartori...