
27
Operation
Example: Analysis with Specified Drying Time
The specified drying time in the example is 10 minutes .
Heating temperature: 130°C
Start of analysis:
Without stability after the cover is closed
End of analysis:
After 10 minutes
Display mode:
Moisture
Step
Key (or instruction) Display
1 . Switch on the moisture
e
Self-test runs
analyzer
2 . Set the drying parameters
(see “Setting the Drying
Parameters” in the chapter
entitled “Configuration”)
3 . Open the sample chamber and
place an unused sample pan
on the pan support
4 . Tare the sample pan:
or
select TAR
as needed
. . .and confirm
5 . Distribute approx . 2 g sample
evenly on the sample pan
6 . Close the sample chamber
The printout header is printed:
see next page
Содержание MA35
Страница 2: ......
Страница 41: ...41 EU Declaration of Conformity ...