Faults
8 Faults
Problem
Cause
Remedy
No segments appear on
the weight display
− No AC power is available
− Check power supply
The weight readout
shows
“LOW”
− No weighing pan on
the scale
− Position the weighing
pan
The weight readout
shows
“HIGH”
− Weighing capacity
exceeded
− Unload the scale
The weight readout
changes constantly
− Unstable ambient
conditions
− Excessive vibration or
draft
− Weighing pan is being
affected at some point
by outside influences
− Set up scale in
another area
− Adjust the scale
settings (see Chap-
ter “3.4 Menu List,”
page 12)
The weight readout is
obviously incorrect
− The sample is not stable
− Scale not tared before
weighing
− Tare before weighing
No weight value is shown
and the padlock symbol
H
is displayed
− PC connection to the
scale has been inter-
rupted activating the
“LOCK” function
− Manual “LOCK” function
is active
− Access the menu to
make the necessary
adjustment and switch
off the “LOCK”
function
− Check the connection
− Switch off the manual
“LOCK” function