Entris
®
II Advanced Line Operating Instructions
53
Technical Data
14 .12 .4 Models BCA12201 | BCA10201 | BCA8201
BCA12201I-
1x
|
BCA12201-
1x
BCA10201I-
1x
|
BCA10201-
1x
BCA8201-
1x
BCA8201I-
1x
Unit
Value
Value
Value
Value
Readability | Scale interval (d)
mg
100
100
100
100
Maximum capacity (Max)
g
12,200
10,200
8,200
8,200
Repeatability
At 5%, typical value
mg
50
50
50
50
At approx . the maximum load, typical value
mg
100
100
100
100
Linearity deviation
Limits
± mg
100
100
300
100
Typical value
± mg
60
60
100
60
Sensitivity drift b10°C and +30°C
±
ppm/K
4
4
7
4
Tare maximum capacity (subtractive)
<100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change
K
2
2
2
Time interval
h
6
6
6
Only for models with approval:
Accuracy class
II
II
II
II
Type
BC-AG
BC-AG
BC-AI
BC-AG
Verification scale interval (e)
mg
1,000
1,000
1,000
1,000
Minimum load (Min)
mg
5,000
5,000
5,000
5,000
Minimum initial weight according to USP (United
States Pharmacopeia), Chap . 41
Optimum minimum initial weight
g
82
82
82
82
Typical stabilization time
100
100
100
100
Typical measurement time
s
≤1.0
≤1.0
≤1.0
≤1.0
Typical stabilization time
s
≤0.9
≤0.9
≤0.9
≤0.9
Recommended calibration weight
External calibrated test weight
g
10,000
10,000
5,000
5,000
Accuracy class in accordance with OIML R111-1
F2
F2
F2
F2