Start up of cassettes
2.4 Integrity Testing
For multiple cassette integrity testing see ”#1 Technical Data Package Integrity
testing Sartocon
®
Cassettes“.
Ensure the filter cassettes are sufficiently wet before measuring the
air diffusion.
2.4.1 Manual Integrity Test
Proceed as follows:
1. Attach a hose to the feed inlet (V
1
), and connect the other end of the hose to
a compressed air source.
2. Close the retentate outlet (V
2
).
3. Adjust the inlet test pressure p
inlet
to the value specified (see
”Specifications“
),
and keep this pressure constant.
4. Close the lower permeate outlet valve V
3
.
5. At a constant inlet pressure pinlet, wait four minutes until the system
has stabilized.
6. Use a flow meter or an inverted burette filled with water to measure the air
diffusion at the upper permeate outlet.
Fig. 1: System set up for manual integrity testing
V
4
Retentate valve
outlet closed
Compressed air
Feed
Inlet
V
2
P
2
P
1
P
3
V
3
Permeate
closed
Cassette
V
1
Permeate
open