manualshive.com logo in svg
background image

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

MPR-214F
 

Pharmaceutical Refrigerator 
with Freezer

 

INSTRUCTION MANUAL

 

Содержание mpr-214f - Commercial Solutions Refrigerator

Страница 1: ...MPR 214F Pharmaceutical Refrigerator with Freezer INSTRUCTION MANUAL ...

Страница 2: ...ll rights reserved Printed in Japan CONTENTS PRECAUTIONS FOR SAFE OPERATION P 2 CAUTIONS FOR USAGE P 6 ENVIRONMENTAL CONDITIONS P 8 REFRIGERATOR COMPONENTS P 9 INSTALLAITON P 12 START UP OF UNIT P 13 OPERATING INSTRUCTIONS P 14 ALARMS AND SAFETY FUNCTIONS P 18 SETTING OF ALARM RESUME TIME P 19 DEFROSTING P 20 MAINTENANCE P 21 AUTOMATIC TEMPERATURE RECORDER P 22 INSTALLATION OF TEMPERATURE RECORDER...

Страница 3: ...t in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symbol shows this symbol means caution this symbol means an action is prohibited this symbol means an instruction must be followed Be sure to keep this manual in a place accessible to users of this unit ...

Страница 4: ...t from the power supply plug before inserting in a power source A dusty plug or improper insertion may pose a hazard Use a power supply outlet with ground earth to prevent electric shock If the power supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cau...

Страница 5: ...sult in fire or injury due to a malfunction Make the power supply to the unit is disconnected when the lamp is replaced as this will prevent electric shock Disconnect the power supply plug if there is something wrong with the unit Continued abnormal operation may cause electric shock or fire If the unit is to be stored unused in an unsupervised area for an extended period ensure that children do n...

Страница 6: ...ot climb onto the unit or do not put articles on the unit This may cause injury by tipping or damage to the unit Do not store bottle or cans in the freezer This may cause injury by broken of containers due to the frozen Do not touch any stock especially metal objects in the freezer with a wet hand This may cause frostbite Always hold the handle when closing the door This will reduce the likelihood...

Страница 7: ...ve a space between the walls of the cabinet and the contents to allow air circulation Do not place items on the floor of the chamber Always close the door firmly The door check lamp is lit when the door is open The alarm buzzer sounds two minutes after door opening The buzzer can be canceled automatically when the door is closed Always open and close the door gently Rough operation may lead fall d...

Страница 8: ... cloth to eliminate the moisture Do not clean the unit with scrubbing brushes acid thinner solvents powdered soap cleanser or hot water These agents can scratch the paint or cause it to peel Plastic and rubber parts can be easily damaged by these materials Especially never use any volatile solvent to clean the plastic or rubber parts If condensation forms on the front glass or frame surface wipe i...

Страница 9: ...umidity 80 for temperature up to 31o C decreasing linearly to 50 relative humidity at 40o C 5 Mains supply voltage fluctuations not to exceed 10 of the nominal voltage 6 Other supply voltage fluctuations as stated by the manufacturer 7 Transient overvoltages according to Installation Categories Overvoltage Categories II For mains supply the minimum and normal category is II 8 Pollution degree 2 in...

Страница 10: ...9 REFRIGERATOR COMPONENTS MPR 214F 3 1 Freezer 2 4 5 10 6 7 8 9 9 4 11 13 13 15 14 Refrigerator 16 ...

Страница 11: ...upper of the fan 8 Lock Turn key clockwise through 180 degree to lock the door The right side lock is for the right upper and lower doors and left side lock is for the left upper and lower doors 9 Access port This port allows cables to be passed into the cabinet 10 Remote alarm terminal rear This is used to alarm the abnormality to the remote location See page 19 for the details 11 Leveling feet U...

Страница 12: ...move Key lock is activated by pressing this key for more than 5 seconds in the temperature display mode See Key lock operation on page 15 5 Numerical value shift key Pressing this key in the setting mode causes the numerical value to change ON OFF of key lock can be selected by pressing this key in the key lock mode 6 Temperature display This indicator shows the chamber temperature set temperature...

Страница 13: ...t in injury from the unit falling or tipping over CAUTION Select a level and sturdy floor for installation This precaution will prevent the unit from tipping Improper installation may result in water spillage or injury from the unit tipping over 5 A location not prone to high humidity WARNING Do not use the unit outdoors Current leakage or electric shock may result if the unit is exposed to rain w...

Страница 14: ...pe or chain 4 Ground earth WARNING Use a power supply outlet with ground earth to prevent electric shock If the power supply outlet is not grounded it is necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit START UP OF UNIT The following p...

Страница 15: ...sing the key shifts up the figure of the current digit 5 Press SET key SET The value is stored in memory and the current refrigerator temperature is displayed 6 Select freezer FREEZ by pressing display select key REF FREEZ The freezer indicator lights and the current freezer temperature is displayed 7 Press SET key SET The current set temperature is displayed and the second digit of the temperatur...

Страница 16: ...played Note Key lock can be set any time when the current refrigerator or freezer temperature is displayed Defrosting of freezer Table 3 below shows the procedure for defrosting of freezer or stopping the freezer operation Table 3 Operation Key operated Indication after operation 1 Select freezer FREEZ by pressing display select key REF FREEZ The current freezer temperature is displayed 2 Press th...

Страница 17: ...ivated when the chamber temperature deviates from set temperature by more than 3o C Note The temperature alarm setting at the factory is as follows High temp alarm Low temp alarm Refrigerator Set temp 5o C Set temp 5o C Freezer Set temp 10o C Set temp 10o C Table 4 Operation Key operated Indication after operation 1 The current refrigerator temperature is displayed 2 Press and hold the numerical v...

Страница 18: ... key Note 1 The first digit of the temperature display blinks 4 Press the set key SET The current set value is displayed and the first digit of the temperature display flashes Pressing the key leads the change of flashed digit 5 Set to 03 by using digit shift key and numerical value shift key Pressing the key shifts up the figure of the current digit 6 Press the set key SET The value is stored in ...

Страница 19: ...r is open Door check lamp is lit Intermittent tone after a delay of 2 minutes Auto return If a key operation is not performed for about 90 seconds in each setting mode Chamber temperature is displayed Setting mode is canceled Key lock When the key lock is ON L1 Key input is unable If the thermal sensor of refrigerator goes open E01 or short circuit E02 Alarm lamp flashes E01 02 and chamber temp ar...

Страница 20: ...g the key shifts up the figure of the current digit 4 Press the set key SET The current set value is displayed and the second digit of the temperature display flashes 5 Set to 020 by using numerical value shift key Pressing the key shifts up the figure in the second digit 6 Press the set key SET The value is stored in memory and the current chamber temperature is displayed The settable alarm resum...

Страница 21: ...tal temperature display Once the forced defrost cycle is complete normal operation resumes The chamber air temperature rises up to about 10o C during the forced defrost cycle Freezer Natural defrost by stopping operation The defrosting is performed by stopping the freezer operation When the frost is built up on the freezer wall defrost the freezer by the following procedure as the freezer has no a...

Страница 22: ...oughly with a cloth soaked in clean water Then wipe the cabinet or accessories unit with a dry cloth to eliminate the moisture Never pour water onto or into the unit Doing so can damage the electrical insulation and may cause electric shock or short circuit The compressor and other mechanical part are completely sealed This unit requires absolutely no lubrication Replacement of lamp Follow the pro...

Страница 23: ...losed with the recorder MTR 0620LH For refrigerator Lower left of front panel In the case of no use of MTR 4014LH MPR S30 MTR 4014LH For freezer Lower left of front panel MPR S30 MTR G3504 For both refrigerator and freezer Lower left of front panel MPR S7 WARING Always disconnect the power supply to the unit before starting the installation of recorder to prevent electric shock or injury Recorder ...

Страница 24: ...sor to the chamber through the port Then replace the thermal insulation and caps both outside and inside Fig 3 Note Make a cut on the caps for access port to pass the capillary tube as shown in the figure below 4 Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the large clips provided as an accessory Fig 4 5 Arrange the capillary tube so that it cannot ...

Страница 25: ...er by pushing the shaft on both sides outward Fig 5 2 Attach the temperature recorder MTR 0620LH to the recorder fixture MPR S30 optional component by referring to the instruction manual enclosed with the recorder fixture Fig 6 3 Remove the cap on the port for taking out the recorder sensor and take out the recorder sensor Then set the recorder fitting to the space of temperature recorder Fig 7 Fi...

Страница 26: ...ps for access port to pass the capillary tube as shown in the figure below 5 Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the large clips enclosed with the recorder fitting MPR S30 optional component Fig 9 6 Arrange the capillary tube so that it cannot contact the shelf Care should be taken for the following points to avoid the crush of the tube when...

Страница 27: ...y pushing the shaft on both sides outward Fig 10 2 Attach the temperature recorder MTR 4014LH to the recorder fixture MPR S30 optional component by referring to the instruction manual enclosed with the recorder fixture Fig 11 3 Remove the cap on the port for taking out the recorder sensor and take out the recorder sensor Then set the recorder fitting to the space of temperature recorder Fig 12 Fig...

Страница 28: ...pillary tube as shown in the figure below 5 Fix the recorder sensor to the attachment hole on the left side of the freezer by using the middle size clips enclosed with the recorder fitting MPR S30 optional component Fig 14 6 Arrange the capillary tube so that it cannot contact the shelf Care should be taken for the following points to avoid the crush of the tube when bending the capillary tube Do ...

Страница 29: ...R G3504 to the recorder fixture MPR S7 optional component by referring to the instruction manual enclosed with the recorder fixture Fig 11 3 Remove the connector cover on the unit by pushing the lock under the cover Connect the power supply connector of the recorder with the connector on the unit Fig 16 4 Remove the cap on the port for taking out the recorder sensor and take out the recorder senso...

Страница 30: ...ote The identification label is adhered to the lead wire to distinguish between the sensor for refrigerator and for freezer Note Make a cut on the caps for access port to pass the lead wire as shown in the figure below 6 Fix the recorder sensor to the attachment hole on the back side of the refrigerator by using the clips enclosed with the recorder fitting MPR S7 optional component Fig 9 7 Fix the...

Страница 31: ...as changed 3 The containers of high temperature load were put in the unit In these cases alarm is removed automatically by running the unit for several hours When the unit does not get cold enough 1 The air exhaust vent is blocked by refrigerator contents 2 A large amount of warm product was put in the unit 3 There is any heat sources in the unit 4 The door is opened frequently 5 The ambient tempe...

Страница 32: ...eet type Freezer Condenser Wire and tube Refrigerator Rear side condenser Freezer Refrigerant R 134a Refrigerator Freezer Defrosting Cycle defrost and forced defrost Refrigerator Natural defrost by stopping operation Freezer Defrost heater 46 3 W Refrigerator Temperature controller Microprocessor control system Temperature display Digital display Thermal sensor Thermister sensor Refrigerator Freez...

Страница 33: ...C No load Noise level 38 dB A scale Maximum pressure 1 3 MPa Refrigerator 1 7 MPa Freezer Rated voltage AC 110 V AC 115 V AC 220 V AC 220 V AC 220 230 240 V Rated frequency 60 Hz 60 Hz 50 Hz 60 Hz 50 Hz Power consumption 160 W 160 W 145 W 155 W 160 W Note The unit with CE mark complies with EC directives 89 336 EEC 93 68 EEC and 73 23 EEC ...

Страница 34: ...it interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance of the unit a The unit is safe to work on b There is some danger see below Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name Model Serial number Date of installation Pharmaceutical Refrigerator MPR 214F with F...

Страница 35: ...7FB6P101349001 1 Oct 2003 Recycled paper SANYO Electric Biomedical Co Ltd 10 15 Hongo 3 Chome Bunkyoku Tokyo 113 8434 Japan Printed in Japan ...