DIGITAL LEAKAGE CLAMP METER

INSTRUCTION MANUAL

 MEMO 

[1] SAFETY PRECAUTIONS

 

*Before use, read the following safety precautions.

 

This instruction manual explains how to use 
your digital leakage clamp meter DLC460F. 
Before use, please read this manual thoroughly 
to ensure correct and safe use. After reading it, 
keep it together with the product for reference 
to it when necessary. 
Using the product in a manner not specified in 
this manual may cause damage to the 
protection function of the product. 
The instructions given under the headings of                   
    WARNING and     CAUTION must be 
followed to prevent accidental 

burn

 and 

electric shock

1-1 Explanation of Warning Symbols

The meanings of the symbols used in this manual 
and attached to the product are as follows:

: Very important instructions for safe use.

The warning messages are intended to 
prevent accidents to operating personnel 
such as burn and electric shock.
The caution messages are intended to 
prevent incorrect handling which may 
damage the product. 

: High voltage hazard

: Ground

: Alternating current (AC) 
: Direct current (DC)
: Resistance

: Buzzer 

: Backlight
: Double insulation or reinforced insulation

1-2 Warning Messages for Safe Use

 

The following instructions are intended to prevent 
personal injury such as burn and electric shock. 
Be sure to follow them when using the meter. 

1.  Never use the meter for the power lines  

exceeding 600VAC to ground.

2.  Voltages over 70VDC or 33Vrms AC (46.7V 

peak) are hazardous to human body. Take 
care so as not to touch them.

3.  Never input signals exceeding the maximum 

rated input value (see 1 3). 

4.  Never use the meter near equipment which 

generates strong electromagnetic waves or is 
charged. 

5.  Never use the meter if the meter or test leads 

are damaged or broken. 

6.  Never use the meter with the case or battery 

lid removed.

7. During measurement, keep your fingers 

behind the finger guard of test leads and the 
meter of the test leads.

8.  To start measurement, do not change the 

meter to another function or range nor replace 
the plugs to other terminals. 

9.  During measurement, do not change function 

switch of the meter nor replace the plugs to 
other terminals.

10. Before starting measurement, make sure that 

the function and range are properly set. 

11. Never use the meter when it is wet or with wet 

hands. 

12. Be sure to use the specified type of test leads.
13. Never attempt repair or modification, except 

for battery replacement. 

14. Always conduct start up inspection and check 

the meter at least once a year. 

15. This meter is for indoor use only. 

1-3 Overload Protection

The maximum rated input value and overload 
protection have been established for the input 
terminals of each function.

[2] APPLICATIONS AND FEATURES

2-1 Applications

This is a digital leakage clampmeter designed for 
measurement in the ranges of CAT. III 600V. This 
meter is useful for measuring leakage or load 
current on power lines and equipment instrument.

2-2 Features

Safety design in compliance with the IEC61010 1. 
Leakage current measurement Resolution 0.01mA
Low Pass Filter function
MAX MIN Hold Function
Backlight Function
Data Hold, Auto Power Save Functions

Classification of overvoltage measurement 
Overvoltage measurement classification 
(CAT. I):

 Line on the secondary side on the 

inside of equipment via a transformer, etc. from 
the receptacle.

Overvoltage measurement classification (CAT. 
II): 

Line on the primary side of equipment with 

power cord to be connected to the receptacle.

Overvoltage measurement classification 
(CAT. III):

 Line from the primary side or branch 

of equipment which directly takes in electricity 
from a distribution board to the receptacle.

Overvoltage measurement classification 
(CAT. IV):

 Line from the service conductor to 

the distribution board.

[3] NAMES OF COMPONENT UNITS

3-1 Main unit

3-2 Display

3-3 Test Lead

In case of test pins cover attached: CAT.III 600V
In case of test pins cover removed: CAT.II 1000V

[4] DESCRIPTION OF FUNCTION

4-1 Power Switch and function switch

Turn this switch to turn on and off the power 
and select a measuring function.

4-2 Date Hold function

When the HOLD button is pressed, the reading 
indicated will be held with      on the display. 
The indicated reading will not change if the 
input signal is changed. When this button is 
pressed again, the function will be disabled and 
the meter will return to the measurement mode 
without      on the display.

4-3 LPF function 

 (mA/ACA)

When the LPF button is pressed,            turns 
on the display. This function cuts current value 
of high frequency. When this button is pressed 
again, the function will be disabled and the 
meter will return to the normal measurement 
mode without        on the display.

Remarks:

Cut off frequency:  3 dB at 180 Hz

4-4  MAX/MIN function

When the MAX/MIN button is pressed, the 
meter enter MAX/MIN mode, fix a measuring 
range and turn                on. Press the button, 
to read  MAX, MIN, current reading sequence.  
Press the button for 1 sec. or more to exit the 
MAX/MIN mode.

Remarks:

Function changes or functional operations will 
cancel the function.

4-5 BACKLIGHT function

When the backlight button is pressed, the 
backlight will be turned on. To disable the 
function, the button is pressed again. The 
backlight will be automatically turned off about 
30 seconds after it was turned on.

4-6 AUTO POWER SAVE function

The meter will go to AUTO POWER SAVE 
function to save battery life about 30 minutes 
after last function switch or button operation. To 
wake up the meter, press any button. To disable 
the function, press any button.

Remarks:

Even in the AUTO POWER SAVE mode, the 
tiny power is still consumed. When the meter is 
not going to be used for an extended period of 
time, be sure to turn off the power switch.
To disable the function, turn the function switch 
to position other than OFF while holding the 
HOLD button.    on the display is turned off 
when Auto Power Save function is disabled.

4-7 Low Battery indication

When the built in batteries have been 
discharged and the voltage has dropped to 
below about 2.2V,    appears in the display. 
When the mark flickers or lights, replace both 
two batteries with new ones.

[5] MEASURING PROCEDURE

 

1. Do not apply an input signal exceeding the 

maximum rated input of each function. 

2. During measurement, do not change the 

function switch.

3. During measurement, keep your finger behind 

finger guard of the test lead and the meter. 

4. When measurement has been finished, remove 

the test leads and CT from the object measured 
and turn the function switch to OFF position.

5-1 Start-up Inspection

Always conduct the start up inspection to ensure 
safety.

1. Be sure that when the meter is turned on, the 

low battery indication       is not flickering or lit. 
If it is flickering or lit, replace the battery with a 
new ones.

2. Do not use the meter if the meter or test lead is 

damaged or broken.

3. Make sure the test leads are not cut.
4. Make sure the meter and your hand are not 

wet.

5-2 Current Measurement (mA)

Remove the test leads from the measuring 
terminals to avoid electric shock.

and

DCV

ACV

ACA

Function

Input
Terminal

Max. Rated
Input Value

Max. Overload
Protection

Do not input
 voltage

DC/AC
600V

AC600 m A

DC/AC
660 V

AC450 A

CT

Data Hold

LPF mode

Auto Power Save

mode

MAX MIN Hold

Units

DC

measurement

AC

measurement

Minus polarity

Digit and 

decimal point

Low battery indication

 1 

 2 

 3 

 4 

 5 

Test pins

Test probe(red)

Finger guards

Test probe(black)

Plugs

Removal test pins cover

SANWA ELECTRIC INSTRUMENT CO., LTD

DempabBldg, 4-4 Sotokanda 2-Chome Chiyoda-ku, Tokyo, Japan

01 1204 5001 6011

Remarks:

·

Clamp the conductor (cable) to measure at the 
center of the trans core (CT).

·

To measure zero phase current, clamp only 
earth wire or all of wires like 2 core or 3 core 
wire together.

·

The meter may malfunction in places where a 
strong magnetic field is present.

·

Do not apply voltage and current at the same 
time.

ACmA current measurement

ACmA

AC100 A

AC400 A

ACmA current measurement

Zero-phase current
clamp earth wire

Zero-phase current
clamp all of wires

ACmA current measurement

Load current
clamp one wire