63
Electromagnetic Compatibility Information
Table 1
Guidance and manufacturer’s declaration – electromagnetic emissions
The SBM 36 is intended for use in the electromagnetic environment specified below. The customer or the user of the SBM 36 should assure that
it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The SBM 36 uses RF energy only for its internal function. Therefore, its RF emis-
sions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class B
The SBM 36 is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The SBM 36 is intended for use in the electromagnetic environment specified below. The customer or the user of the SBM 36 should assure that
it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels charac-
teristic of a typical location in a typical commercial or hospital
environment.
Содержание SBM 36
Страница 66: ...66...
Страница 67: ...67...
Страница 68: ...68 751 380 0514 Irrtum und nderungen vorbehalten...