Instructions for use
v2
Histolog® Scanner v2 003 Histolog® Dish 004
1223
version 1.11_EN (2022-05-26)
Страница 1: ...Instructions for use v2 Histolog Scanner v2 003 Histolog Dish 004 1223 version 1 11_EN 2022 05 26...
Страница 2: ...ght and all rights are reserved No part of this document may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopying and recording for any purpose othe...
Страница 3: ...1 5 2019 06 21 Removed 11 Improve image quality for print Add description of symbol for quantity in package 1 6 2020 08 31 Update SamanTree Medical s address 1 7 2020 11 16 Update list of compatible c...
Страница 4: ...3 Other accessories 12 7 3 Operating principle 12 8 Use of the Histolog Scanner 13 8 1 Before use 13 8 1 1 Installing in use location 13 8 1 2 Switching ON 13 8 1 3 Mounting Histolog Dish 13 8 2 Afte...
Страница 5: ...est procedure 21 14 2 Troubleshooting 22 14 3 Export logs 24 15 Product catalog codes 25 16 Device specifications technical information 26 16 1 Technical specifications 26 16 1 1 Histolog Scanner 26 1...
Страница 6: ...w the local regulations for separate collection of Electrical and Electronic Equipment CE marking of conformity with Directive 98 79 EC and Regulation EU 2017 746 UKCA marking of conformity with the M...
Страница 7: ...d these instructions for use before setting up and using the device the first time Keep the instructions for use accessible for future reference Refer to the Histolog Dip instructions for use for info...
Страница 8: ...ad with mass above 1 kg Do not operate the Histolog Scanner in case of damage to the imaging window G The protective earthing connection of the Histolog Scanner is performed through the power cord P c...
Страница 9: ...adiation in the visible wavelength range is emitted from the imaging window G during imaging The Histolog Scanner embeds a Class IIIB laser component with the following characteristics Wavelength 488...
Страница 10: ...s The Histolog Dish a single use protection preventing direct contact of the tissue specimen with the Histolog Scanner during imaging The Histolog Dip fluorescent histological stain The Histolog Scann...
Страница 11: ...thernet port only for maintenance purposes under guidance of SamanTree Medical authorized service personnel F Orientation mark G Imaging window H Handles I Lid J Swivel caster wheels with brakes K App...
Страница 12: ...rage devices with security lock features such as PIN protection are not compatible with the Histolog Scanner Refer to the Histolog Scanner Software Application instructions for use for detailed instru...
Страница 13: ...istolog Scanner is performed through the power cord P connection If this cable is damaged the device must not be connected to the mains power Only use a power cord P having adequate ratings for the Hi...
Страница 14: ...do not want to shutdown the Histolog Scanner Power OFF the Histolog Scanner device 3 Wait 15 seconds or until the No Signal message is displayed on the monitor and then switch OFF the power switch C...
Страница 15: ...al interface of the Histolog Dish with the surface of interest pointing downwards NOTE Specimen size should not exceed that of the Histolog Dish see 16 1 Furthermore the tissue specimen should be posi...
Страница 16: ...ble with the Histolog Scanner Active compound Tested with Alcohol Ethanol 71 Propanol 50 Quaternary ammonia compounds DDAC Didecyldimethylammonium chloride CAS 7173 51 5 0 075 Benzalkonium Chloride be...
Страница 17: ...H and not other parts of the device for pushing pulling Move the device no faster than walking speed Move the device over a flat surface Avoid surfaces and obstacles that could hamper the normal rolli...
Страница 18: ...The device can then be moved to the storage area When in place lock the four wheel brakes J in downward position to prevent unintended movement of the Histolog Scanner See 16 1 for specified environme...
Страница 19: ...ware Application Update of Histolog Scanner Software Application may be installed only by SamanTree Medical authorized service personnel 12 2 Fuse replacement Fuses are located in a compartment above...
Страница 20: ...ling biomedical waste 13 2 Histolog Scanner Follow the local regulations for separate collection of electrical and electronic equipment 13 3 Histolog Dip Dispose of the Histolog Dip in accordance with...
Страница 21: ...s than a minute to complete 3 Once the test procedure is completed the status of the different components is displayed in a message window to the bottom right corner of the screen OK means that the co...
Страница 22: ...ront surface of the touchscreen monitor A and let it dry completely Press and hold the Touch Disabled button for 3 seconds to enable the touch function Touchscreen has entered in an abnormal state of...
Страница 23: ...ntact with the optical interface of the Histolog Dish Reposition the tissue specimen to ensure better contact Use tools and or weights to help in positioning and improving the contact between the tiss...
Страница 24: ...the logs 1 Connect a USB storage device to the USB port D NOTE The USB storage device should have at least 1 GB of available storage capacity 2 Tap the Histolog Scanner button in the upper left corner...
Страница 25: ...anner appliance power inlet T6 3 AH 250VAC 5mm 20mm PN446 Power Cord Swiss plug PN458 Power Cord Euro plug PN482 Power Cord United Kingdom plug PN460 Power Cord North America plug Documentation Docume...
Страница 26: ...5 W Power Cord Length 5 m USB port standard 3 0 Ethernet port connector RJ45 Fuse type T 6 3AL 250VAC 5x20mm slow blow Display Active screen size diagonal 547 mm 21 5 Active screen size W x H 476 64 m...
Страница 27: ...2010 Temperature 15 35 C Humidity 30 80 non condensing Pressure 80 106 kPa Transport Conditions Temperature 0 60 C Humidity 10 80 non condensing Pressure 70 106 kPa Storage Conditions Temperature 0 60...
Страница 28: ...fferential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines...
Страница 29: ...ths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment...
Страница 30: ...ufacturer NOTE At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE These guidelines may not apply in all situations Electromagnetic propagation is affected by abs...