
Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
31
3.3.3 < TC >: Setting the Time Constants
The range of time constant is from 10 us to 3000 s. Adjust the value through the left and right buttons and the spin
knob.
The greater the time constant, the smaller the ENBW, the longer the system response and the higher the
measurement accuracy.
Table 2. Time Constants
10 μs
3 ms
1 s
300 s
30 μs
10 ms
3 s
1000 s
100 μs
30 ms
10 s
3000 s
300 μs
100 ms
30 s
1 ms
300 ms
100 s
3.3.4 < Filter>: Setting the Roll off Slope of the Low Pass Filter
<6dB/oct>
The roll off slope is 6 dB/oct.
<12dB/oct>
The roll off slope is 12 dB/oct.
<18dB/oct>
The roll off slope is 18 dB/oct.
<24dB/oct>
The roll off slope is 24 dB/oct.
At the same measurement accuracy, a greater roll off slope can reduce the time constant, making a faster response.
3.3.5 < Synchronous >: Setting the Status of the Synchronous Filter
<OFF>
Close the synchronous filter.
<200Hz>
Open the synchronous filter. It’s useful at a low reference frequency (< 20 Hz). When the input signal
frequency is very low, the low pass filter can’t get stable results or it needs a long time to do that. At
this time, synchronous filter helps.
3.4 [DISPLAY]
Press [DISPLAY] button and enter the submenu as is shown in Fig.19. It displays the results of measurement,
including the values of <R>, <
θ
>, <X> and <Y>.