Figure 8.9: You can drag the leads up, down or switch sides.
8.7
Choose the stimulation electrodes automatically
1.
Stop stimulation
if stimulation is running.
2.
Change the Electrode Selection Mode icon (see Figure 8.2) until the Automatic
Electrode Selection icon
is displayed.
3.
In the electrode display window, touch the electrode that you wish to use as the
stimulation electrode
.
4.
Touch the stimulation electrode again to cycle through electrode configurations.
Tripolar: 1 lead
Bipolar: 1 lead
Tripolar: 2 leads
Bipolar: 2 leads
5.
The CPA will automatically select the stimulation
, return
,
measurement
and reference
electrodes for you.
You may steer the current between stimulation electrodes (see Section 9.4).
You can also manually change the electrode configuration (see Section 9.3).
6.
Touch Set Defaults icons in the Settings menu, Stimulation Settings menu, and ECAP
Measurement Settings menu (see Figure 8.10) to set standard stimulation
parameters.
7.
Restart stimulation on the new electrodes.
Note: You cannot change the stimulation or return electrodes while stimulation is running.