Intra-Operative Testing
Place lead(s) based on required anatomical location.
Connect the lead to the lead adapter
Figure 10.1: Lead adapter kit.
1.
Place the tip of the proximal connector end of the lead into the end of the lead
adapter slot. The stylet fits into the groove on the lead adapter
(Note: it is not necessary to remove the lead Stylet)
Figure 10.2: Placing tip of proximal connector into lead adapter slot with stylet fitting in
groove.
2.
Push the lead down into slot completely using finger (see Figure 10.3), so that the
lead is flush with the top of the slot. The lead should not move when fully pushed
into the slot. After the lead is in the slot, press down again along its length to ensure
it is secure in the slot.