Инструкция и гарантия для продукта Salter BPA-9200 доступны для загрузки бесплатно на manualshive.com. Скачайте руководство пользователя на русском языке, чтобы узнать все подробности о функциях и гарантийных условиях этого удивительного продукта.
HoMedics Group Ltd
HoMedics House, Somerhill Business Park, Five Oak Green Road, Tonbridge, Kent TN11 0GP, UK.
www.salterhousewares.co.uk
IB-BPA-9200-GB-0314-01
E
Register your product today at:
www.homedicsgroup.com/register
0120
Страница 1: ...BPA 9200 BLOOD PRESSURE MONITOR Instructions and Guarantee E BLOOD PRESSURE MONITOR Instructions and Guarantee E ...
Страница 2: ...ld be taken in packing the blood pressure monitor so that it is not damaged while in transit This undertaking is in addition to a consumer s statutory rights and does not affect those rights in any way For UK Sales and Service contact HoMedics Group Ltd HoMedics House Somerhill Business Park Five Oak Green Road Tonbridge Kent TN11 0GP UK Helpline Tel No 01732 360783 e mail support homedics co uk F...
Страница 3: ...ff is wrapped around the upper arm The cuff circumference is limited to 22cm 48cm approx 8 21 32 18 29 32 CONTRAINDICATION It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer CONTENTS AND DISPLAY INDICATORS LCD display Cuff Memory Indicator Date Time Display Alternating Systolic Pressure MEM button START button Diastolic Pressure Pulse Rate Display Altern...
Страница 4: ... 1997 A1 2005 A2 2009 Non invasive sphygmomanometers Part 3 Supplementary requirements for electro mechanical blood pressure measuring systems SPECIFICATIONS Product name Blood Pressure Monitor Model BPA 9200 GB Classification Internally powered Type BF applied part IPX0 No AP or APG Continuous operation Machine size Approx 138mm 98mm 48mm 5 7 16 x 3 27 32 x 1 7 8 Cuff circumference 30cm 42cm 11 1...
Страница 5: ...e please see part ELECTROMAGNETIC COMPATIBILITY INFORMATION 12 If Irregular Heartbeat IHB brought by common arrhythmias is detected in the procedure of blood pressure measurement a signal of will be displayed Under this condition the Electronic Sphygmomanometers can keep function but the results may not be accurate it s suggested that you consult with your physician for accurate assessment There a...
Страница 6: ...ttery or turn off the monitor it will enter Clock Mode and LCD will display time and date by turns See picture 2 2 1 Picture 2 Picture 2 1 Picture 2 2 b While the monitor is in Clock Mode pressing both the START and MEM button simultaneously a beep is heard and the month will blink at first See picture 2 2 Press the button START repeatedly the day hour and minute will blink in turn While the numbe...
Страница 7: ...Never dry the cuff in clothes dryer or iron it Clean the cuff after the usage of every 200 times is recommended 5 BODY POSTURE DURING MEASUREMENT Sitting Comfortably Measurement a Be seated with your feet flat on the floor and don t cross your legs b Place palm upside in front of you on a flat surface such as a desk or table c The middle of the cuff should be at the level of the right atrium of th...
Страница 8: ...a health care professional for interpretation of pressure measurements 7 DISPLAYING STORED RESULTS a After the measurement you can review the measurements in the current memory bank by pressing button MEM Now the LCD displays the amount of the results in the current bank See picture 7 Picture 7 Picture 7 1 Picture 7 2 b Alternatively press MEM button in Clock Mode to display the stored results The...
Страница 9: ...ton displays the respective results measured previously Picture 7 6 Picture 7 7 Picture 7 8 Picture 7 9 e When displaying the stored results the monitor will turn off automatically after 1 minute of no operation You can also press the button START to turn off the monitor manually 8 DELETING MEASUREMENTS FROM THE MEMORY When any result is displaying keeping on pressing button MEM for three seconds ...
Страница 10: ...olated the instructions The technical alarm condition outside the rated range is preset in the factory and cannot be adjusted or inactivated This alarm condition is assigned as low priority according to IEC 60601 1 8 The technical alarm is non latching and need no reset The signal displayed on LCD will disappear automatically after about 8 seconds 11 TROUBLESHOOTING 1 PROBLEM POSSIBLE CAUSE SOLUTI...
Страница 11: ...tion Do not immerse the monitor in water as this will result in damage to the monitor 03 If this monitor is stored near freezing allow it to acclimate to room temperature before use 04 Do not attempt to disassemble this monitor 05 If you do not use the monitor for a long time please remove the batteries 06 It is recommended the performance should be checked every 2 years or after repair Please con...
Страница 12: ...es com servicecentres EXPLANATION OF SYMBOLS ON UNIT Symbol for THE OPERATION GUIDE MUST BE READ The sign background colour blue The sign graphical symbol white Symbol for WARNING Symbol for TYPE BF APPLIED PARTS The cuff is type BF applied part Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check wit...
Страница 13: ...y network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Table 2 For all ME EQUIPMENT and ME SYSTEMS Guidance and manufacturer s declaration electromagnetic immunity The BPA 9200 is intended for use in the electromagnetic environment specified below The customer or the user of th...
Страница 14: ... transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflec...
Страница 15: ...tput power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation...
Страница 16: ...oup Ltd HoMedics House Somerhill Business Park Five Oak Green Road Tonbridge Kent TN11 0GP UK www salterhousewares co uk IB BPA 9200 GB 0314 01 E Register your product today at www homedicsgroup com register 0120 ...