Salter AIRE Elite

 

 

Compressor   

Instruction Manual 

Part No: 8350-1 / 8352-1 / 8353-1   

      

 

 

Thank you for selecting the 

Salter AIRE

 

Elite

 Compressor.  

Salter Labs is an innovative, industry-leading manufacturer 

of respiratory care devices.    Please contact your local Salter 

Labs dealer for information about additional products. 

 

SAVE THESE INSTRUCTIONS

READ ALL INSTRUCTIONS BEFORE USE.

 

 

Contents 

Important  Safeguards………………………………….………………1 

Introduction…………………………………………………………………2 

Product Description…………………….……………………………..3 

Operation……………………………………………………………..…….4 

Cleaning……………………………………………..………………………..5 

Storage……………………………………………………………………….6 

Maintenance………………………………………………………………7 

Expected Service Life…………………………………………………8 

Troubleshooting…………………………………………………………9 

Specifications.……………………………….……………………………10 

Accessories….……………………………………………………………11 

Appendix A………………………………………………………………12

 

 
1. Important Safeguards 

NOTE, CAUTION, WARNING, AND SYMBOLS: 

Important information is highlighted by using the following: 

 

NOTE

 Indicates information that user should pay special attention 

to. 

CAUTION

 Indicates correct operating or maintenance procedures in 

order to prevent damage to or destruction of the equipment or other 

property. 

WARNING

 Indicates potential danger that requires correct 

procedures or practicesin order to prevent personal injury

 

Symbols: 

Off, disconnection from the mains 

On, connection to the mains 

 

Alternating Current (AC) 

IP2X 

Protected against solid foreign objects having a 

diameter of 12.5 mm and greater. No protection against 

vertically falling water drops, Keep dry ! 

 

Attention 

 

Class II 

 

 

Refer to instruction manual/ booklet 

 

“BF” symbol, indicate this product is according to the 

degree of protecting against electric shock for the type 

BF equipment. 

 

Temperature limitation 

 

Disposal of Electrical & Electronic Equipment (WEEE):   
This product should be handed over to an applicable 

collection point for the recycling of electrical and 

electronic equipment. For more detailed information 

about the recycling of this product, please contact your 

local city office, household waste disposal service or the 

retail store where you purchased this product.

 

 

CAUTION

 

U.S. Federal Law restricts this device to sale by or on the 

order of a physician. 

WARNING

 To reduce the risk of burns, electrocution, fire or 

injury to persons: 

1.

 

Always unplug this product immediately after using. 

2.

 

Do not use while bathing, showering, dish washing, or close to 

water sources of any kind. 

3.

 

Do not place or store product where it can fall or be pulled 

into a tub or sink. 

4.

 

Do not place in or drop into water or other liquid. 

5.

 

Do not reach for a product that has fallen into water. Unplug 

immediately. 

6.

 

This product should never be left unattended when plugged in. 

7.

 

Close supervision is necessary when this product is used by, 

on or near children or invalids. Choking accident may result 

from a child swallowing a small part that has become detached 

from the device or its accessories. 

8.

 

Use this product only for its intended use as described in this 

manual. Use this product only under doctor’s direction. Do not 

use attachments not recommended by the manufacturer. 

9.

 

Never operate this product if a) it has a damaged cord or plug, 

b) it is not working properly, c) it has been dropped or 

damaged, d) it has been dropped into water. Return the 

product to a specified service center for examination and 

repair. 

10.

 

Keep the cord away from heated surfaces. 

11.

 

Never block the air openings of this product or allow objects to 

fall or be inserted into the air vent openings or place it on a 

soft surface such as bed or couch, where the air openings may 

be blocked. 

12.

 

Never use while sleeping or feeling drowsy. 

13.

 

Never drop or insert any object into any opening or hose. 

14.

 

No modification of this equipment is allowed. 

15.

 

Do not modify this equipment without authorization of the 

manufacturer. 

16.

 

If this equipment is modified, appropriate inspection and 

testing must be conducted to ensure continued safe use of the 

equipment. 

17.

 

Do not use in outdoors or operate where aerosol (spray) 

products are being used or where oxygen is being 

administered in a closed environment such as an oxygen 

reservoir. 

18.

 

Do not wrap the power cord around the compressor (main 

unit).   

19.

 

Disconnect the power plug by pulling the plug, not by pulling 

on the compressor (main unit), or the cord. 

20.

 

If the power cord or plug becomes frayed or otherwise 

damaged, do not use.   

21.

 

Do not place heavy objects on the power cord, or bend and 

pull the cord harder than necessary. These actions could cause 

an electric shock or fire. 

22.

 

Potential allergic reactions to accessible materials used in the 

Compressor Nebulizer equipment. If any signs of allergic 

reaction or hypersensitivity happen, stop the treatment 

immediately, and notify the doctor or nurse. 

23.

 

Potential contact injuries for patients used in the Compressor 

Nebulizer equipment. If any contact injuries happen, stop the 

treatment, and notify the doctor or nurse. 

WARNING

 EMC Statement 

This equipment has been tested and found to comply with the 

limits for medical devices to the EN 60601-1:2007. These limits are 

designed to provide reasonable protection against harmful 

interference in a typical medical installation. This equipment 

generates uses and can radiate radio frequency energy and, if not 

installed and used in accordance with the instructions, may cause 

harmful interference to other devices in the vicinity. However, 

there is no guarantee that interference will not occur in a particular 

installation. If this equipment does cause harmful interference to 

other devices, which can be determined by turning the equipment 

off and on, the user is encouraged to try to correct the interference 

by one or more of the following measures: 

 

Reorient or relocate the receiving device. 

 

Increase the separation between the equipment. 

 

Connect the equipment into an outlet on a circuit different from 

that      to which the other device(s) are connected.

 

 

Consult the manufacturer or field service technician for help.

 

CAUTION

 If there is a possibility of electro-magnetic interference 

with mobile phones, please increase the distance (3.3m) between 

devices or turn off the mobile phone. 

2. Introduction 

2.1 Intended Use 

The 

Salter AIRE

 

Elite

 Compressor System is intended to provide a 

source of compressed air for aerosol therapy. It is used in 

conjunction with a jet (pneumatic) nebulizer to produce medicated 

aerosols for inhalation by pediatric and adult patients with 

respiratory symptoms. 

CAUTION

 Aerosolize liquid medication except Pentamidine for 

inhalation by the patients. Indications for therapy include asthma, 

chronic bronchitis, infection of the upper respiratory tract, chronic 

obstructive pulmonary disease (COPD) and other respiratory 

disorders in accordance with a medical doctor’s prescription. Except 

the usage mentioned above, please do not use this product for any 

other purpose. This device can be used with adults or pediatric 

patients under physician’s prescription.

   

2.2 Safety Precaution Instruction 

When using this electrical product, especially when children are 

present, one should always follow basic safety precautions. Do not 

install, maintain or operate this equipment without reading, 

understanding and following the proper 

Salter AIRE 

Elite

 

Compressor System instruction manual, otherwise injury or 

damage may result.   

For 120V only-

 This appliance has a polarized plug (one blade is 

wider than the other). To reduce the risk of electric shock, this plug 

is intended to fit into a polarized outlet only one way. If the plug 

does not fit fully into the outlet, reverse the plug.  If it still does 

not fit, contact a qualified electrician. Do not modify the plug in any 

way. 

3. Product Description 

1. Power Switch 

5. Power Cord 

 

2. Filter Cap (disposable filter inside)  6. Nebulizer Cup Holder 

3. Air-Outlet Connector 

7. Cooling Air Openings 

4. Integrated Carrying Handle 

 

4. Operation 

NOTE

 Before initial operation, the nebulizer cup assembly should 

be cleaned following instructions described in the “Cleaning” 

section. 

WARNING

 Before connecting the power cord, make sure the I/O 

(ON/OFF) switch is in the O (OFF) position. 

 

The plug is also served to disconnect the device. Do not 

position the equipment so that it is difficult to operate 

the disconnecting device. 

4.1 Daily Use Operation 

CAUTION

 The Salter AIRE Elite Compressor System is designed for 

intermittent use only. Do not operate it continuously for more than 

30 minutes for a single use without turning it off and following a 

cooling period for least 30 minutes. 

1.

 

Before each use inspect the 

Salter AIRE

 

Elite

 

Compressor 

and nebulizer cup assembly for damage or wear, replace as 

needed. 

2.

 

Place the 

Salter AIRE

 

Elite

 Compressor on a table or other 

flat stable surface. Be sure you can easily reach the controls 

when seated. Do not use this device on the floor. 

3.

 

With the power switch in the O (OFF) position, plug the power 

cord into an appropriate electrical wall outlet. 

4.

 

Connect one end of the tubing to the compressor air-outlet 

connector. 

5.

 

Assemble the nebulizer cup and add indicated medication to 

the nebulizer’s cup before use. 

6.

 

Attach the other end of the tubing into the air-inlet connector 

founded at the bottom of the nebulizer cup. 

7.

 

Turn on the 

Salter AIRE

 

Elite

 Compressor by pressing the 

power switch to the I (ON) position and begin treatment. 

8.

 

If treatment needs to be interrupted, simply press power 

switch to O (OFF) position. 

9.

 

When the treatment is complete, turn off the compressor by 

pressing the power switch to O (OFF) position and unplug the 

unit from the electrical outlet. 

Equipment is not suitable for use in the presence of a 

flammable anesthetic mixture with air or with oxygen or 

nitrous oxide. This device has no AP or APG protection.

 

CAUTION

 During the treatment, the patient should not touch the 

outer case due to expected

 

rise in unit temperature.   

4.2 Safety Overload 

NOTE

 Do not exceed 30 minutes of continuous operation.

 

1.

 

The motor of this device has a built-in thermal overload 

protector. Should the motor overheat, the protector will 

automatically shut off the motor. Should this occur, turn the 

I/O (ON/OFF) switch to the O (OFF) position and allow the 

motor to cool down for approximately 30 minutes before turn 

it on again. 

2.

 

If the overload protector shuts off the motor frequently, you 

may have an unstable voltage situation. 

3.

 

If the unit shuts down and cannot restart, it may need to be 

replaced. Call your provider immediately.

 

5. Cleaning 

5.1 Compressor Outer Case Cleaning 

WARNING

 Electric shock hazard. Do not remove outer case of 

this unit. All disassembly and maintenance of this unit must be 

done by a qualified service technician. Refer servicing to 

qualified service personnel. 

 

WARNING

 This unit does not require oil. Do not attempt to 

lubricate any internal parts. 

WARNING

 Unplug unit before cleaning. Do not submerge in 

water for cleaning. 

1.

 

Wipe the main unit with a damp cloth every few days to keep 

it dust-free. 

2.

 

Do not use any powdered type cleaners or soap. Do not 

submerge the unit into water. 

 

5.2. Nebulizer Cup Cleaning 

Clean the nebulizer after each use.    Refer to the cleaning 

instructions supplied with your nebulizer 

800987 Rev D (Side 1) 

  Printed on 15 April 2015 