Quick Reference Guide 

Tissue-Tek® Film® 

Sakura Finetek Europe B.V. | Flemingweg 10a, 2408 AV Alphen a/d Rijn | NL 

Phone: +31 (0) 88 592 00 00 | Fax: +31 (0) 88 592 00 01 | [email protected] | www.sakura.eu  

Loading station

 

Xylene 

bottle

 

Waste 

bottle

 

Film 

loading  

area

 

Film door

 

Cover

 

Power switch

 

Control panel

 

Loading 

door

 

 

Start Up 

The Tissue-Tek

®

 Film™ automated coverslipper is 

designed to coverslip biological specimens that are 
mounted on standard 25 x 75 mm (1 x 3 inch) 
microscope slides. The use of a special, resin-
coated, film and xylene eliminates the need for 
coverglass and mounting media.  

 

Preparation before coverslipping 

 
Hardware 
Placement of loading station
 
1.

 

Insert the loading station and fill it with xylene. 

 
Placement of xylene and waste bottle 
1.

 

Fill the xylene bottle and insert this in the 
instrument. 

2.

 

Insert the waste bottle. 

 
Placement of the coverslipping film 
1.

 

Load the coverslipping film. 

2.

 

Instructions for loading the coverslipping film 
can be found on the diagram located on the 
back wall of the instrument.  

3.

 

The film is wound through positioning spools 
and rollers until it has reached the 
coverslipping area.  

4.

 

Press “film” to lead the film automatically 
trough roller seven. 

 
 

 

Содержание Tissue-Tek Film

Страница 1: ...e use of a special resin coated film and xylene eliminates the need for coverglass and mounting media Preparation before coverslipping Hardware Placement of loading station 1 Insert the loading station and fill it with xylene Placement of xylene and waste bottle 1 Fill the xylene bottle and insert this in the instrument 2 Insert the waste bottle Placement of the coverslipping film 1 Load the cover...

Страница 2: ...he priming level can be selected in a range of 1 5 3 being the default setting 2 Film Length refers to the length of the film placed on the slide to cover the specimen There are four selectable lengths 45mm 50mm 55mm and 60mm 3 Extended Prime is required each time the xylene bottle is refilled This prime ensures that there are no bubbles in the priming line The extended prime will prime the line f...

Страница 3: ... the setting item 3 Moves the selection item upward or leftward Holding the button will cause continuous action 4 Moves the selection item downward or rightward Holding the button will cause continuous action 5 ENTER Saves the setting item and returns to the previous screen 6 FILM Feeds the film 7 PRIME Conducts priming with the prime corresponding to the dispensing level 1 5 To prevent air bubble...

Страница 4: ... operating manual section 4 1 for more detailed information Most common method for removal 1 Place the slides in acetone until film can be peeled off approx 5 min to 10 min 2 Place the slides in clean acetone for 5 min 3 Place the slides in ethanol 100 4 Place the slides in xylene 5 Slides are now ready for further treatment Maintenance see operating manual section 6 for more detailed information ...

Страница 5: ...tied and cleaned 2 Unscrew the thumb screw on the receiver tray 3 Remove the tray 4 Remove film particles from the tray 5 Clean the tray with xylene Yearly Preventive maintenance by a Sakura representative or a qualified service engineer Maintenance task Frequency Exterior cleaning Daily Waste bottle Daily Check xylene bottle Daily Cleaning the slide baskets Weekly Cleaning the loading station Wee...

Отзывы: